Zhou H, Wu Z, Ma L, Wu W, Yang S, Wang Q, Yuan X, Wu L, Lin X, Tan J. Assessing immunologic function through CD4 T-lymphocyte ahenosine triphosphate levels by ImmuKnow assay in Chinese patients following renal transplantation.
Transplant Proc 2012;
43:2574-8. [PMID:
21911125 DOI:
10.1016/j.transproceed.2011.04.012]
[Citation(s) in RCA: 9] [Impact Index Per Article: 0.8] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 02/10/2011] [Revised: 03/10/2011] [Accepted: 04/19/2011] [Indexed: 01/21/2023]
Abstract
OBJECTIVE
Balancing immunosuppression to prevent rejection while minimizing infection/drug toxicity risk is a challenge in organ transplantation. Drug monitoring alone or with functional monitoring is inadequate to measure the immune response after transplantation. The Food and Drug Administration (FDA)-approved immune monitoring assay, ImmuKnow, offers an noninvasive method to assess the immune status of transplanted patients by measuring adenosine triphosphate (ATP) released from CD4 T cells. Herein, we have evaluated ATP levels reflecting the immune responses of Chinese kidney transplant recipients as a monitoring parameter to guide treatment after transplantation.
METHODS
From October 2008 to March 2010, we recruited 259 kidney transplant patients who were divided into four groups: stable (n = 174), postoperative infection (n = 32), postoperative rejection (n = 16), and high-dose corticosteroid treatment (n = 33). The ImmuKnow assay was performed to measure CD4 T-cell ATP levels. No prisoners or organs from prisoners were used in the study.
RESULTS
Receiver operating characteristics measurements indicated an ATP predictive range of 238 to 497 ng/mL to monitor immune responses after transplantation and immunosuppressive therapy. To identify patients with infection, we used a cutoff ATP value of 238 ng/mL with 100% specificity and positive predictive value and 92.9% sensitivity. To identify patients with rejection, we used a value of 497 ng/mL with 91.5% sensitivity. Compared with the 225 to 525 ng/mL ATP levels recommended by the FDA, our target values showed similar or better diagnostic accuracy.
CONCLUSION
We provide additional data to monitor immunosupressant treatment of Chinese kidney transplant patients.
Collapse