Matuszak D, Gaddy GD, Aranovich GL, Donohue MD. Cosorption effect in gas chromatography: flow fluctuations caused by adsorbing carrier gases.
J Chromatogr A 2005;
1063:171-80. [PMID:
15700469 DOI:
10.1016/j.chroma.2004.11.064]
[Citation(s) in RCA: 3] [Impact Index Per Article: 0.2] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/16/2022]
Abstract
Adsorbing carrier gases have a number of advantages in analytical and preparative gas chromatography, such as clearer detector signals and higher column efficiencies. This work shows that adsorbing carrier gases also may be useful because they cause the mobile phase flow rate to become unsteady after injecting a small amount of sample. This work shows that a 100 microL sample of helium can liberate enough carbon dioxide carrier gas from a zeolite 5A packed column at 373 K, that the departure from the steady-state flow rate had an upper lobe area of 586 microL of carrier gas. This was confirmed by coupling a modified Langmuir kinetic model with the Ergun equation.