Fricke K, Konecny F, El-Warrak A, Hodgson C, Cadieux-Pitre H, Hill T, Sobot R. In-vivo characterization of left-ventricle pressure-volume telemetry system in swine model.
Biomed Microdevices 2016;
18:75. [PMID:
27492638 DOI:
10.1007/s10544-016-0094-7]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 10/21/2022]
Abstract
We present in-vivo study related to the use of our implantable RF telemetry system for pressure-volume (PV) cardiac monitoring in a animal subject. We implant a commercial MEMS PV sensor into the subject's heart left-ventricle (LV), while the telemetry system is implanted outside of the heart and connected to the sensor with a 7-microwires tether. The RF telemetry system is suitable for commercial application in medium sized subjects, its total volume of 2.475cm(3) and a weight of 4.0g. Our designed system is 58 % smaller in volume, 44 % in weight and has a 55 % reduction in sampling power over the last reported research in PV telemetry. In-vivo data was captured in both an acute and a freely moving setting over a 24 hour period. We experimentally demonstrated viability of the methodology that includes the surgical procedure and real-time monitoring of the in-vivo data in a freely moving subject. Further improvements in catheter design will improve the data quality and safety of the subject. This real-time implantable technology allows for researchers to quantify cardiac pathologies by extracting real-time pressure-volume loops, wirelessly from within freely moving subjects.