Sokoloff AJ, Yang Z, Sargolzaei S, Strait K, Krasnopeyev A, Easley KA, Mimche S, Ghovanloo M. Magnetic implants in the tongue for assistive technologies: Tests of migration; oromotor function; and tissue response in miniature pigs.
Arch Oral Biol 2017;
81:81-89. [PMID:
28499234 DOI:
10.1016/j.archoralbio.2017.04.034]
[Citation(s) in RCA: 1] [Impact Index Per Article: 0.1] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 09/22/2016] [Accepted: 04/30/2017] [Indexed: 10/19/2022]
Abstract
OBJECTIVE
Uncertain biological consequences of titanium-magnet (Ti-mag) tongue implants constrain application of the Tongue Drive System (TDS), a brain-tongue-computer interface for individuals with severe physical impairment. Here we describe oromotor function and tongue tissue response following Ti-Mag implantation and explantation in the miniature pig, an animal model with a tongue similar in size to humans.
DESIGN
A 1.8×6.2mm Ti-mag tracer was implanted into the anterior tongue in five Yucatan minipigs. X-rays were taken immediately and >six days after implantation to evaluate tracer migration. In three minipigs, the tracer was explanted >16days after implantation. Twenty-five days post-explantation, tongue tissue was harvested and processed for histological and immunohistochemical (IHC) markers of healing. In two minipigs tissue markers of healing were evaluated post-mortem following >12days implantation. Drink cycle rate (DCR) was characterized to determine the impact of procedures on oromotor function.
RESULTS
Neither implantation (N=5) nor explantation (N=3) changed DCR. X-rays revealed minimal tracer migration (N=4, 0-4mm). By histology and IHC a robust capsule was present two weeks post-implantation with limited fibrosis. Explantation produced localized fibrosis and limited muscle remodeling.
CONCLUSIONS
These findings suggest the safety of Ti-mag anterior tongue implants for assistive technologies in humans.
Collapse