Nicia SB, van Veelen TA, Stens J, Koopman MMW, Boer C. Detection of volume loss using the Nexfin device in blood donors.
Anaesthesia 2015;
71:163-70. [PMID:
26566706 DOI:
10.1111/anae.13283]
[Citation(s) in RCA: 3] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Accepted: 07/23/2015] [Indexed: 12/13/2022]
Abstract
We investigated which haemodynamic parameters derived from Nexfin non-invasive continuous arterial blood pressure measurements are optimal to detect controlled volume loss in spontaneously breathing subjects. Haemodynamic monitoring was performed in 40 whole-blood donors. Mean arterial pressure, cardiac index, systemic vascular resistance index and pulse pressure variation were recorded during controlled breathing, and a Valsalva manoeuvre was performed before and after blood donation. Blood donation resulted in a reduction in cardiac index (from 3.96 ± 0.84 l.min(-1) .m(2) to 3.30 ± 0.61 l.min(-1) .m(2) ; p < 0.001), an increase in systemic vascular resistance (from 1811 ± 450 dyn.s.cm(-5) .m(2) to 2137 ± 428 dyn.s.cm(-5) .m(2) ; p < 0.001) and an increase in pulse pressure variation (from 13.4 ± 5.1 to 15.3 ± 5.4%; p = 0.02). The area under the receiver operating characteristic curve to detect volume loss was highest for cardiac index (0.94, 95% CI 0.88-0.99) and systemic vascular resistance (0.90, 95% CI 0.82-0.99). Nexfin is a non-invasive haemodynamic monitor that can feasibly detect volaemic changes in spontaneously breathing subjects.