Ammari WG, Al-Qadhi Z, Khalil M, Tayyem R, Qammaz S, Oriquat G, Basheti IA, Chrystyn H. Indacaterol determination in human urine: validation of a liquid-liquid extraction and liquid chromatography-tandem mass spectrometry analytical method.
J Aerosol Med Pulm Drug Deliv 2014;
28:202-10. [PMID:
25229261 DOI:
10.1089/jamp.2014.1153]
[Citation(s) in RCA: 5] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 10/24/2022] Open
Abstract
BACKGROUND
Indacaterol is a novel once-a-day inhaled ultra-long-acting β2-agonist. Quantitative bioanalysis supports pharmacokinetic and clinical research. The aim of the current work was to validate an in-house developed high performance liquid chromatography (HPLC)-tandem mass spectrometry (MS/MS) analytical method for indacaterol determination in human urine samples.
METHODS
A liquid-liquid extraction method has been developed to extract indacaterol from human urine samples using ethyl acetate. Indacaterol dry extract was reconstituted with 200 μL of the mobile phase (acidified water:methanol (30:70, v/v)) of which 5 μL was needed for the HPLC-MS/MS analysis. Indacaterol was eluted on a reversed C18 stationary phase with an isocratic mobile phase at a flow of 1 mL/min. Formoterol was the internal standard (IS). The MS/MS detection was employed with a turbo-ion spray ionization in the positive ion mode. A consensus of the international Guidelines for Bioanalytical Method Validation was followed.
RESULTS
Indacaterol was detected at a mass to charge ratio (m/z) of 393.3 and its MS/MS daughter at 173.2. The retention times of indacaterol and IS were 1.60 and 1.20 min, respectively. Validated calibration curves were linear over a range of 0.075-100 ng/mL with correlation coefficients (r)≥0.990. The curves' regression weighting factor was 1/x. Method specificity was established in six different human urine batches. No matrix interference was observed. The intra- and inter-batch precision and accuracy within±20% (at lower limit) and±15% (other quality control (QC) levels) were confirmed. The indacaterol mean recovery (precision) percentages at Low, Mid, and High QC levels were 93.5 (3.84), 89.8 (2.15), and 92.2 (2.17), respectively. Short-term, long-term, freeze-thaw, and auto-sampler stability results were accepted.
CONCLUSIONS
A specific, accurate and precise HPLC-MS/MS method has been validated for indacaterol quantification in human urine. This simple method is reproducible and robust to support future, indacaterol-related pharmacokinetic, bioequivalence and clinical studies.
Collapse