van Elsäcker A, Nikolic T, Scheffer MG, van Gelder BM. Increase of ventricular output inducing ventricular afterpotential sensing and ventricular safety pacing in a biventricular implanted cardioverter defibrillator.
Europace 2009;
12:142-5. [PMID:
19801565 DOI:
10.1093/europace/eup283]
[Citation(s) in RCA: 2] [Impact Index Per Article: 0.1] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/13/2022] Open
Abstract
Programming maximum right ventricular output in a patient with a biventricular implanted cardioverter defibrillator resulted in ventricular oversensing and ventricular safety pacing in the same cardiac cycle.