1
|
Kreimer F, Gotzmann M. Pacemaker-induced atrial fibrillation reconsidered-associations with different pacing sites and prevention approaches. Front Cardiovasc Med 2024; 11:1412283. [PMID: 38957332 PMCID: PMC11217490 DOI: 10.3389/fcvm.2024.1412283] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 04/04/2024] [Accepted: 06/05/2024] [Indexed: 07/04/2024] Open
Abstract
The incidence of atrial fibrillation (AF) is significantly higher in patients with pacemakers than in the general population, which could be due to patient characteristics and the diagnostic tool of the pacemaker in detecting atrial high-rate episodes and subclinical AF, but also to the pacemaker itself providing AF-promoting conditions. It is well known that high ventricular pacemaker burden increases the likelihood of AF occurrence. However, the sites of atrial and ventricular pacing may also influence the risk for AF. The conventional sites for atrial and ventricular pacing are in the right atrial appendage and in the right ventricular apex. However, growing evidence suggests that alternative pacing sites may be superior for the prevention of AF. Bachmann bundle pacing, for example, promotes interatrial excitation conduction, resulting in atrial synchronicity and a shorter total atrial activation time, which may be preventive for the occurrence of AF. Moreover, in recent years, new ventricular pacing sites have come into focus with His bundle and left bundle branch pacing. In addition to the hemodynamic and electrophysiological cardiac benefits, these new options may also offer benefits in the prevention of AF. This review provides an overview of pacing-induced AF mechanisms and the association with different pacing sites, as well as approaches for prevention of pacing-induced AF, highlighting different sites and modes of atrial pacing and the newer sites of ventricular pacing.
Collapse
Affiliation(s)
- Fabienne Kreimer
- Department of Cardiology and Rhythmology, St Josef Hospital Bochum, University Hospital of the Ruhr University Bochum, Bochum, Germany
| | | |
Collapse
|
2
|
Osswald B. [Surgical basics of cardiac implantable electronic device implantation from skin incision through closure]. Herzschrittmacherther Elektrophysiol 2022; 33:476-482. [PMID: 36066608 DOI: 10.1007/s00399-022-00892-6] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 07/07/2022] [Accepted: 08/04/2022] [Indexed: 06/15/2023]
Abstract
Implantation of pacemakers is generally considered a "minor intervention". Younger colleagues obtain their skills from experienced ones; ideally not just over the course of one or two interventions under surveillance but until a certain level of confidence is achieved. In Germany, certification is still optional. The German Cardiology Society (DSC, "Deutsche Gesellschaft für Kardiologie") provides expertise courses to obtain basic knowledge of pacemaker therapy; the GCS/GSCVS has been offering certification modules since 2013 to acquire fundamental knowledge from experts and basic technical skills by simulator training. The present article illustrates the major aspects of pacemaker implantation procedures and some avoidable pitfalls.
Collapse
Affiliation(s)
- Brigitte Osswald
- Elektrophysiologisch Chirurgie, Medizinische Klinik I, Johanniter-Krankenhaus Duisburg-Rheinhausen, Kreuzacker 1-7, 47228, Duisburg, Deutschland.
| |
Collapse
|
3
|
Cobb DB, Gold MR. The Role of Atrioventricular and Interventricular Optimization for Cardiac Resynchronization Therapy. Heart Fail Clin 2017; 13:209-223. [DOI: 10.1016/j.hfc.2016.07.017] [Citation(s) in RCA: 6] [Impact Index Per Article: 0.9] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Indexed: 01/30/2023]
|
4
|
Interatrial septal pacing to suppress atrial fibrillation in patients with dual chamber pacemakers: A meta-analysis of randomized, controlled trials. Int J Cardiol 2016; 219:421-7. [DOI: 10.1016/j.ijcard.2016.06.093] [Citation(s) in RCA: 10] [Impact Index Per Article: 1.3] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 12/20/2015] [Revised: 05/23/2016] [Accepted: 06/21/2016] [Indexed: 11/23/2022]
|
5
|
Cobb DB, Gold MR. The Role of Atrioventricular and Interventricular Optimization for Cardiac Resynchronization Therapy. Card Electrophysiol Clin 2015; 7:765-779. [PMID: 26596818 DOI: 10.1016/j.ccep.2015.08.008] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 06/05/2023]
Abstract
Many patients with left ventricular systolic dysfunction may benefit from cardiac resynchronization therapy; however, approximately 30% of patients do not experience significant clinical improvement with this treatment. AV and VV delay optimization techniques have included echocardiography, device-based algorithms, and several other novel noninvasive techniques. Using these techniques to optimize device settings has been shown to improve hemodynamic function acutely; however, the long-term clinical benefit is limited. In most cases, an empiric AV delay with simultaneous biventricular or left ventricular pacing is adequate. The value of optimization of these intervals in "nonresponders" still requires further investigation.
Collapse
Affiliation(s)
- Daniel B Cobb
- Division of Cardiology, Department of Medicine, Medical University of South Carolina, Charleston, SC, USA
| | - Michael R Gold
- Division of Cardiology, Department of Medicine, Medical University of South Carolina, Charleston, SC, USA.
| |
Collapse
|
6
|
Krychtiuk KA, Nürnberg M, Volker R, Pachinger L, Jarai R, Freynhofer MK, Wojta J, Huber K, Weiss TW. Effects of AV-delay optimization on hemodynamic parameters in patients with VDD pacemakers. Wien Klin Wochenschr 2014; 126:270-7. [PMID: 24652015 DOI: 10.1007/s00508-014-0520-x] [Citation(s) in RCA: 4] [Impact Index Per Article: 0.4] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 04/30/2013] [Accepted: 02/04/2014] [Indexed: 02/05/2023]
Abstract
AIM Atrioventricular (AV) delay optimization improves hemodynamics and clinical parameters in patients treated with cardiac resynchronization therapy and dual-chamber-pacemakers (PM). However, data on optimizing AV delay in patients treated with VDD-PMs are scarce. We, therefore, investigated the acute and chronic effects of AV delay optimization on hemodynamics in patients treated with VDD-PMs due to AV-conduction disturbances. METHODS In this prospective, single-center interventional trial, we included 64 patients (38 men, 26 women, median age: 77 (70-82) years) with implanted VDD-PM. AV-delay optimization was performed using a formula based on the surface electrocardiogram (ECG). Hemodynamic parameters (stroke volume (SV), cardiac output (CO), heart rate (HR), and blood pressure (BP)) were measured at baseline and follow-up after 3 months using impedance cardiography. RESULTS Using an ECG formula for AV-delay optimization, the AV interval was decreased from 180 (180-180) to 75 (75-100) ms. At baseline, AV-delay optimization led to a significant increase of both SV (71.3 ± 15.8 vs. 55.3 ± 12.7 ml, p < 0.001, for optimized AV delay vs. nominal AV interval, respectively) and CO (5.1 ± 1.4 vs. 3.9 ± 1.0 l/min, p < 0.001), while HR and BP remained unchanged. At follow-up, the improvement in CO remained stable (4.9 ± 1.3 l/min, p = 0.09), while SV slightly, but significantly, decreased (to 65.1 ± 17.6, p < 0.01). CONCLUSION AV-delay optimization in patients treated with VDD-PMs exhibits immediate beneficial effects on hemodynamic parameters that are sustained for 3 months.
Collapse
Affiliation(s)
- Konstantin A Krychtiuk
- Department of Internal Medicine II-Division of Cardiology, Medical University of Vienna, Waehringer Guertel 18-20, 1090, Vienna, Austria,
| | | | | | | | | | | | | | | | | |
Collapse
|
7
|
Sonne C, Bott-Flugel L, Hauck S, Hadamitzky M, Lesevic H, Demetz G, Braun D, Wolf P, Hausleiter J, Schömig A, Kolb C. Three-dimensional echocardiographic optimization improves outcome in cardiac resynchronization therapy compared to ECG optimization: a randomized comparison. PACING AND CLINICAL ELECTROPHYSIOLOGY: PACE 2013; 37:312-20. [PMID: 24164640 DOI: 10.1111/pace.12281] [Citation(s) in RCA: 11] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Subscribe] [Scholar Register] [Received: 05/08/2013] [Revised: 07/06/2013] [Accepted: 08/12/2013] [Indexed: 11/30/2022]
Abstract
AIMS There is little consensus on optimal atrioventricular (AV) and ventricular-to-ventricular (VV) intervals in cardiac resynchronization therapy (CRT). The aim of this study was to examine a novel combination of Doppler echocardiography (DE) and three-dimensional echocardiography (3DE) for individualized AV- and VV-interval optimization compared to conventional electrocardiogram (ECG) optimization. METHODS In this double-blind, randomized controlled trial, 77 patients (male: 57, age: 68 ± 10 years) with severely reduced ejection fraction (EF), New York Heart Association (NYHA) class III or IV, and wide QRS complex (>120 ms) have been included. Patients were randomized to either AV- and VV-interval optimization using DE and 3DE (group 1, n = 39) or ECG (group 2, n = 38). 3DE was performed in all patients for the evaluation of left ventricular (LV) dimensions, EF and systolic dyssynchrony index (SDI), and NYHA class obtained before CRT and after 3 months. Primary endpoint of the study was clinical response to CRT, defined as a reduction of NYHA class by ≥1 score. Secondary endpoints were change of EF, LV volumes, and SDI. RESULTS There were significantly more responders in group 1 (82%) than in group 2 (58%, P = 0.021). Similarly, group 1 showed a larger increase in EF (7.0 ± 6.0% vs 3.4 ± 5.6%, P = 0.015) and a more pronounced reduction of SDI (-4.5 ± 5.9% vs -1.5 ± 5.6%, P = 0.039) than group 2. CONCLUSION Compared with conventional ECG optimization, this novel echocardiographic optimization protocol resulted in a significantly higher response rate, improved LV systolic function, and may be used to select the optimal AV and VV intervals in CRT.
Collapse
Affiliation(s)
- Carolin Sonne
- Klinik für Herz- und Kreislauferkrankungen, Deutsches Herzzentrum München, Technische Universität München, Munich, Germany
| | | | | | | | | | | | | | | | | | | | | |
Collapse
|
8
|
Pakarinen S, Toivonen L. Minimizing ventricular pacing by a novel atrioventricular (AV) delay hysteresis algorithm in patients with intact or compromised intrinsic AV conduction and different atrial and ventricular lead locations. Ann Med 2013; 45:438-45. [PMID: 23768003 DOI: 10.3109/07853890.2013.801710] [Citation(s) in RCA: 4] [Impact Index Per Article: 0.4] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Indexed: 11/13/2022] Open
Abstract
PURPOSE To investigate if an advanced AV search hysteresis (AVSH) algorithm, Ventricular Intrinsic Preference (VIP(™)), reduces the incidence of ventricular pacing (VP) in sinus node dysfunction (SND) with both intact and compromised AV conduction and with intermittent AV block regardless of the lead positions in the right atria and the ventricle. METHODS Patients were classified as having intact AV (AVi) conduction if the PR interval was ≤ 210 ms on ECG and 1:1 AV conduction during atrial pacing up to 120 bpm with PR interval ≤ 350 ms. Otherwise the AV conduction was classified as compromised (AVc). Both AVi and AVc patients were randomized to VIP ON or OFF. VIP performed an intrinsic AV conduction search every 30 s for three consecutive atrial cycles with the extension of the sensed and paced AV (SAV/PAV) delays from basic values of 150/200 ms to 300/350 ms. Extended AV intervals were allowed for three cycles when VP occurred before returning to basic AV delays. The primary end-point was %VP at 12 months. RESULTS Among 389 patients, 30.1% had intact and 69.9% had compromised AV conduction. The mean %VP at 12 months was 9.6% by VIP compared to 51.8% with standard AV settings in patients with AVi (P < 0.0001) and 28.0% versus 78.9% (P < 0.0001) with AVc. With VIP, excessive %VP among most used lead positions was not seen. Conversely, when VIP was off %VP was low only in patients who had leads in the RA septal-RV septal position (23.0%). CONCLUSIONS VIP feature reduces VP both in patients with SND and with intermittent heart block regardless of the lead positions in the right atria and the ventricle.
Collapse
Affiliation(s)
- Sami Pakarinen
- Department of Cardiology, Helsinki University Central Hospital, Meilahti Hospital, Helsinki, Finland.
| | | |
Collapse
|
9
|
Kanadaşı M, Caylı M, Sahin DY, Sen Ö, Koç M, Usal A, Batur MK, Demirtaş M. The effect of different atrioventricular delays on left atrium and left atrial appendage function in patients with DDD pacemaker. Echocardiography 2011; 28:626-32. [PMID: 21718356 DOI: 10.1111/j.1540-8175.2011.01404.x] [Citation(s) in RCA: 2] [Impact Index Per Article: 0.2] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/26/2022] Open
Abstract
BACKGROUND Although it has been known that optimization of atrioventricular delay (AVD) has favorable effect on the left ventricular functions in patients with DDD pacemaker, the effect of different AVDs on left atrium (LA) and left atrial appendage (LAA) functions has not been exactly evaluated. The aim of the present study was to assess the effect of different AVDs on LA and LAA functions in DDD pacemaker implanted patients with atrioventricular block. METHODS Forty-eight patients with DDD pacemaker were enrolled into the study. Patients were divided into two groups according to the echocardiographic diastolic function: Group I (normal diastolic function) and Group II (diastolic dysfunction). LAA emptying velocity on pulsed wave Doppler and LAA late systolic wave velocity by using tissue Doppler were recorded. Patients were paced for five successive continuous pacing periods of 10 minutes duration using five selective AVDs (80-250 ms). RESULTS Significant effect on LA and LAA functions has not been observed by the setting of AVD in Group I. However, when the AVD was gradually shortened form 150 ms to 80 ms, LA and LAA functions gradually decreased in Group II patients. When AVD increased to 200 ms, LA and LAA functions were improved. Further increase in AVD resulted in decreased LA and LAA functions. CONCLUSION Setting of AVD has not significant effect on the LA and LAA functions in patients with normal diastolic function, but moderate prolongation of AVD in physiological limits improved LA and LAA functions in DDD pacemaker implanted patients with diastolic dysfunction.
Collapse
Affiliation(s)
- Mehmet Kanadaşı
- Department of Cardiology, Faculty of Medicine, Çukurova University Adana, Turkey
| | | | | | | | | | | | | | | |
Collapse
|
10
|
LEVIN VADIM, RAZAVI MEHDI, COLL ROBERT, COLES Jr JAMESA, SAMBELASHVILI ALEKSANDRE. Interatrial Conduction Correlates with Optimal Atrioventricular Timing in Cardiac Resynchronization Therapy Devices. PACING AND CLINICAL ELECTROPHYSIOLOGY: PACE 2011; 34:443-9. [DOI: 10.1111/j.1540-8159.2010.02988.x] [Citation(s) in RCA: 6] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 11/29/2022]
|
11
|
JONES RCHRISTOPHER, SVINARICH TOM, RUBIN ANDREW, LEVIN VADIM, PHANG ROBERT, MURILLO JAIME, SAMBELASHVILI ALEKSANDRE. Optimal Atrioventricular Delay in CRT Patients Can Be Approximated Using Surface Electrocardiography and Device Electrograms. J Cardiovasc Electrophysiol 2010; 21:1226-32. [DOI: 10.1111/j.1540-8167.2010.01807.x] [Citation(s) in RCA: 27] [Impact Index Per Article: 1.9] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Indexed: 11/29/2022]
|
12
|
Nitardy A, Langreck H, Dietz R, Stockburger M. Reduction of right ventricular pacing in patients with sinus node dysfunction through programming a long atrioventricular delay along with the DDIR mode. Clin Res Cardiol 2008; 98:25-32. [DOI: 10.1007/s00392-008-0716-z] [Citation(s) in RCA: 11] [Impact Index Per Article: 0.7] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 10/16/2007] [Accepted: 08/19/2008] [Indexed: 11/28/2022]
|
13
|
CHIRIFE RAUL, PASTORI JULIO, MOSTO HUGO, ARRASCAITE MÓNICA, SAMBELASHVILI ALEKSANDRE. Prediction of Interatrial and Interventricular Electromechanical Delays from P/QRS Measurements: Value for Pacemaker Timing Optimization. PACING AND CLINICAL ELECTROPHYSIOLOGY: PACE 2008; 31:177-83. [DOI: 10.1111/j.1540-8159.2007.00966.x] [Citation(s) in RCA: 13] [Impact Index Per Article: 0.8] [Reference Citation Analysis] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 11/27/2022]
|
14
|
Kenny D, Walsh KP. Noncatheter-Based Delivery of a Single-Chamber Lumenless Pacing Lead in Small Children. PACING AND CLINICAL ELECTROPHYSIOLOGY: PACE 2007; 30:834-8. [PMID: 17584263 DOI: 10.1111/j.1540-8159.2007.00769.x] [Citation(s) in RCA: 17] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 11/29/2022]
Abstract
OBJECTIVES The model 3830 lead (SelectSecure Medtronic, Minneapolis, MN, USA) is a bipolar, fixed-screw, 4.1-F pacing lead designed for site-selective pacing. Implantation is usually performed using an 8-F deflectable catheter system. This catheter is not ideal for smaller children because of both the sheath size and the relatively large deflected curves. We describe a simpler noncatheter-based delivery system in seven children. METHODS A 4.1-F SelectSecure lead was introduced via a 5-F SafeSheath (Thomas Medical, Malvern, PA, USA) placed in the left subclavian vein. The SelectSecure lead was passed into the inferior vena cava (IVC) and a loop created, which was then withdrawn into the right atrium. Once in position, the lead was screwed into the myocardium, the SafeSheath was peeled off, and the lead connected to the generator. RESULTS From March 2005 until September 2006, five right atrial leads and two right ventricular leads were implanted in seven patients (six female) with a median weight of 15 kg (8.1-19.4). All leads were successfully placed with excellent pacing thresholds. The median screening time was 7.1 minutes (4.8-11.4) with a median radiation dose of 83 cGycm(2). There were no procedural complications and no lead displacements seen on a median follow-up of 10 months. CONCLUSIONS Delivery of the 4.1-F SelectSecure pacing lead to the right heart is possible using a noncatheter-based delivery system. This is effective and safe and does not require the use of a larger delivery system. This allows these thin isodiametric pacing leads to be used advantageously in small children.
Collapse
Affiliation(s)
- Damien Kenny
- Cardiac Department, Our Lady's Hospital for Sick Children, Crumlin, Dublin, Ireland.
| | | |
Collapse
|
15
|
Optimization of atrioventricular delay by surface electrocardiography during dual chamber pacing. Chin Med J (Engl) 2006. [DOI: 10.1097/00029330-200603020-00004] [Citation(s) in RCA: 6] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/26/2022] Open
|