Hennessy-Strahs S, Bermudez CA, Acker MA, Bartoli CR. Toward a Standard Practice to Quantify von Willebrand Factor Degradation During Left Ventricular Assist Device Support.
Ann Thorac Surg 2020;
112:1257-1264. [PMID:
33227272 DOI:
10.1016/j.athoracsur.2020.09.039]
[Citation(s) in RCA: 6] [Impact Index Per Article: 1.5] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 03/28/2020] [Revised: 08/24/2020] [Accepted: 09/20/2020] [Indexed: 02/01/2023]
Abstract
BACKGROUND
Continuous-flow left ventricular assist devices (LVADs) cause degradation of von Willebrand factor (VWF) multimers and bleeding. Multiple techniques exist to characterize VWF deficiency. However, a standard methodology has not been established in LVAD patients. Toward this goal, we evaluated 4 methods to quantify VWF multimers.
METHODS
We collected paired blood samples from patients (n = 48) before and after 1 week of LVAD support. After 652 ± 59 days of support, patients were classified as bleeders (≥1 bleeding episode) or nonbleeders. VWF multimers were resolved with electrophoresis and immunoblotting, the gold-standard to evaluate VWF multimers. We evaluated 4 quantification methods.
RESULTS
Each method demonstrated significant VWF degradation during LVAD support vs a paired, pre-LVAD sample (method 1, VWF length: 48 of 48 patients, -10% ± 1%, P < .0001; method 2, VWF density: 40 of 48, -34% (interquartile range, -46% to -8%), P < .0001; method 3, pre-LVAD to LVAD ratio: 46 of 48, 17 ± 5: 10 ± 1, P < .0001; method 4, LVAD/pre-LVAD index: 46 of 48, 57% (interquartile range, 50% to 73%), P < .0001). Bleeding occurred in 27 of 48 patients. Method 1 demonstrated significantly fewer VWF multimers in bleeders compared with nonbleeders (-11% ± 1% vs -8% ± 1%; P = .01). Other methods did not demonstrate this potentially important clinical relationship.
CONCLUSIONS
A standardized methodology is needed to quantify VWF multimer degradation with mechanical circulatory support devices. Novel method 1 successfully quantified the patient-specific change in VWF multimer length during LVAD support and demonstrated a difference in VWF multimers between bleeders and nonbleeders. Adoption of consensus methodology will assist to standardize patient-specific bleeding risk, inform anticoagulation and antiplatelet therapy, and evaluate LVAD hemocompatibility.
Collapse