Martinez JG, De Sousa J, Dompnier A, Martins-Oliveira M, Israel CW, Teijeira E, Rubin JM, Sebag F, Martino M, Michel Y, Marques P. Efficacy and safety of novel left ventricular pacing leads: 1-year analysis of the NAVIGATOR trial.
Open Heart 2024;
11:e002517. [PMID:
38316493 PMCID:
PMC10860098 DOI:
10.1136/openhrt-2023-002517]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 10/13/2023] [Accepted: 01/03/2024] [Indexed: 02/07/2024] Open
Abstract
OBJECTIVES
Assess safety and performance of novel quadripolar preshaped left ventricular (LV) leads: NAVIGO 4LV 2D ('S shaped') and NAVIGO 4LV ARC ('U shaped').
METHODS
Patients indicated for cardiac resynchronisation therapy were enrolled in a multicentre, prospective, controlled study (NAVIGATOR, NCT03279484). Patients were implanted with either a NAVIGO 4LV 2D or ARC lead, and assessed at 10 weeks, 6, 12 and 24 months post-implant. Co-primary safety and performance endpoints were assessed at 10 weeks. Safety endpoint was the patients' rate free from lead-related complications. Performance endpoint was the rate of patients with successful lead performance, defined as LV pacing threshold ≤2.5 V at 0.5 ms on at least one pacing vector, and the absence of phrenic nerve stimulation at the final programmed configuration. Lead-related complications and electrical parameters were monitored throughout study.
RESULTS
A NAVIGO 4LV lead was successfully implanted in 211 out of 217 patients (97.2%). The safety endpoint was met, with 100% and 96.1% of patients free from complications for NAVIGO 4LV 2D and ARC, respectively. The performance endpoint was met with 98.1% and 98.9% of patients with a successful lead performance for NAVIGO 4LV 2D and ARC, respectively. Over 12 months, the global complication-free rate for both leads was 97.1% (95% CI: 93.71% to 98.70%), with a mean pacing capture threshold of 1.23 V±0.73 V and a mean impedance of 951 Ω±300.1 Ω.
CONCLUSION
A high implantation success rate and low complication rate was reported for the novel NAVIGO 4LV 2D and ARC leads, along with successful performance up to 12 months.