Jaiyeola C, Chen AY, Kalarickal PL, Bhatt BP. An Unexpected Magnet Response of a Biotronik Pacemaker in Automatic Mode: A Case Report.
A A Pract 2022;
16:e01617. [PMID:
36067419 DOI:
10.1213/xaa.0000000000001617]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/05/2022]
Abstract
In patients with cardiovascular implantable electronic devices, asynchronous pacing is necessary for surgeries with a risk of electromagnetic interference. Magnets are often used for asynchronous pacing. In this case report, magnet application to a Biotronik Evia DR-T pacemaker (Biotronik) programmed to a magnet response mode called automatic magnet mode (AUTO) led to a brief period of asynchronous pacing before reverting to prior settings, an unexpected response that was not revealed on the preoperative interrogation report. For Biotronik pacemakers programmed to the AUTO mode, changes in programming are needed for asynchronous pacing.