Liang X, Beltran CJ, Liu C, Shen J, Li H, Furutani KM. Technical note: Delivery benefit and dosimetric implication of synchrotron-based proton pencil beam scanning using continuous scanning mode.
Med Phys 2023;
50:5252-5261. [PMID:
37115647 DOI:
10.1002/mp.16434]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 12/10/2022] [Revised: 03/03/2023] [Accepted: 04/13/2023] [Indexed: 04/29/2023] Open
Abstract
BACKGROUND
Discrete spot scanning (DSS) is the commonly used method for proton pencil beam scanning (PBS). There is lack of data on the dose-driven continuous scanning (DDCS).
PURPOSE
To investigate delivery benefits and dosimetric implications of DDCS versus DSS for PBS systems.
METHODS
The irradiation duty factor, beam delivery time (BDT), and dose deviation were simulated for eight treatment plans in prostate, head and neck, liver, and lung, with both conventional fractionation and hypofractionation schemes. DDCS results were compared with those of DSS.
RESULTS
The DDCS irradiation duty factor (range, 11%-41%) was appreciably improved compared to DSS delivery (range, 4%-14%), within which, hypofractionation schemes had greater improvement than conventional fractionation. With decreasing stop ratio constraints, the DDCS BDT reduction was greater, but dose deviation also increased. With stop ratio constraints of 2, 1, 0.5, and 0, DDCS BDT reduction reached to 6%, 10%, 12%, and 15%, respectively, and dose deviation reached to 0.6%, 1.7%, 3.0%, and 5.2% root mean square error in PTV DVH, respectively. The 3%/2-mm gamma passing rate was greater than 99% with stop ratio constraints of 2 and 1, and greater than 95% with a stop ratio of 0.5. When the stop ratio constraint was removed, five of the eight treatment plans had a 3%/2-mm gamma passing rate greater than 95%, and the other three plans had a 3%/2-mm gamma passing rate between 90% and 95%.
CONCLUSIONS
The irradiation duty factor was considerably improved with DDCS. Smaller stop ratio constraints led to shorter BDTs, but with the cost of larger dose deviations. Our finding suggested that a stop ratio of 1 constraint seems to yield acceptable DDCS dose deviation.