He C, Pant A, Le AH. Impact of the Acuros XB spatial discretization error on ArcCHECK VMAT QA for small-field SBRT.
J Appl Clin Med Phys 2023;
24:e14100. [PMID:
37563870 PMCID:
PMC10476976 DOI:
10.1002/acm2.14100]
[Citation(s) in RCA: 3] [Impact Index Per Article: 3.0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 02/03/2023] [Revised: 04/26/2023] [Accepted: 06/27/2023] [Indexed: 08/12/2023] Open
Abstract
PURPOSE
To evaluate the impact of the Acuros XB spatial discretization errors on ArcCHECK volumetric modulated arc therapy (VMAT) QA for small-field SBRT plans.
METHODS
Eighteen SBRT VMAT arcs that failed the ArcCHECK VMAT QA were retrospectively analyzed. Plan verification doses were calculated using Eclipse Acuros XB, and absolute 3%/2 mm gamma passing rates were calculated to compare ArcCHECK and MapCHECK2 with MapPHAN. Verification doses were recalculated using AAA in Eclipse and with the EGSnrc Monte Carlo package. In addition, error-reduced Acuros XB doses were calculated by subdividing the entire arc into several sub-arcs ("split-arc" method), with the angular ranges of the sub-arcs optimized to balance accuracy and efficiency. Relative gamma passing rates were calculated and compared for the four methods: (1) Acuros XB; (2) AAA; (3) EGSnrc Monte Carlo; and (4) the split-arc method.
RESULTS
The absolute gamma passing rates were below 90% for ArcCHECK and above 95% for MapCHECK2. The averaged relative gamma passing rates were (1) 84.7% for clinical Acuros XB; (2) 96.8% for AAA; (3) 98.8% for EGSnrc Monte Carlo; and (4) 96.8% for the split-arc method with 60° sub-arc angle. Compared to the clinical Acuros XB, the split-arc method improved the relative gamma passing rate by 12.1% on average. No significant difference was found between AAA and the split-arc method (p > 0.05).
CONCLUSION
The Acuros XB spatial discretization errors can significantly impact the ArcCHECK VMAT QA results for small-field SBRT plans. The split-arc method may be used to improve the VMAT QA results.