1
|
Esmaelbeigi A, Kalinowski J, Tomic N, Rivard MJ, Vuong T, Devic S, Enger SA. E-Brachy: New dosimetry package for electronic brachytherapy sources. Med Phys 2024. [PMID: 39460996 DOI: 10.1002/mp.17462] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 02/28/2024] [Revised: 08/06/2024] [Accepted: 09/18/2024] [Indexed: 10/28/2024] Open
Abstract
BACKGROUND Large reported variability in the material composition and geometrical components of the Xoft electronic high-dose-rate brachytherapy Causes inter-source discrepancy in the source output. This variability is due to the manual manufacturing and assembly of the sources. PURPOSE This study aimed to develop a dosimetry software tool called E-Brachy to characterize the Xoft source and quantify the discrepancies in its photon spectrum and dosimetric properties. METHODS E-Brachy is based on the Geant4 Monte Carlo toolkit and consists of two parts. In part one, the geometry and material composition for the source received in the computer-aided design format from the vendor were converted to the geometry description markup language format using the GUIMesh Python tool and integrated into the E-Brachy software. There was a large variation in material composition and thickness for some of the tube components. The simulation started from electrons and resulted in x-ray generations in the anode region. Multithreading, a track length estimation, and the uniform bremsstrahlung splitting variance reduction techniques were used to decrease the simulation time and increase the x-ray production. The photon energy, position, and momentum were saved into a phase space file as the photon exited the source, but before interacting with the external environment. The obtained x-ray energy spectrum was compared with measurements from the National Institute of Standards and Technology (NIST). In part two, by sampling from the generated photons, the dose rates and dosimetric parameters according to the TG-43 protocol were calculated for model S7500 and compared to the ones previously calculated for model S700 source, which were deemed identical by the manufacturer. RESULTS The material composition that resulted in the most similar spectrum as the measured NIST spectrum with Pearson's correlation coefficient of 0.99 and a calculated Euclidean difference of0.061 ± 0.001 $0.061\,\pm \,0.001$ keV was chosen for further dosimetric analysis of the model S7500 source. Characteristic peaks showed the presence of tungsten, yttrium, and silver in the source components. Differences in dose rates between the two source models surpassed 20% for polar anglesθ ≥ 150 ∘ $\theta \,\ge \,150^\circ$ , reaching a peak atr = 3 $r\,=\,3$ cm andθ = 175 ∘ $\theta \,=\,175^\circ$ . The differences in the radial dose function values were within 5%. The relative difference in percentage between the anisotropy function values of the two models was closer to 0 for smaller θ $\theta$ values, but at higher polar angles, they increased to 300%. CONCLUSIONS A software package called E-Brachy was successfully developed for the characterization and dosimetry of Xoft electronic brachytherapy sources. E-Brachy can be combined with spectral measurements to investigate the inter- and intra-source variability. The software package was tested by comparing the simulated spectra from the S7500 Xoft source model with NIST measurements and its TG-43 parameters with the S700 model. The TG-43 parameters between the two sources significantly exceed the recommendations of TG-56.
Collapse
Affiliation(s)
- Azin Esmaelbeigi
- Medical Physics Unit, Department of Oncology, Faculty of Medicine, McGill University, Montreal, Quebec, Canada
| | - Jonathan Kalinowski
- Medical Physics Unit, Department of Oncology, Faculty of Medicine, McGill University, Montreal, Quebec, Canada
| | - Nada Tomic
- Medical Physics Unit, Department of Oncology, Faculty of Medicine, McGill University, Montreal, Quebec, Canada
| | - Mark J Rivard
- Department of Radiation Oncology, Alpert Medical School of Brown University, Providence, Rhode Island, USA
| | - Te Vuong
- Department of Radiation Oncology, Jewish General Hospital, Montreal, Quebec, Canada
| | - Slobodan Devic
- Department of Radiation Oncology, Jewish General Hospital, Montreal, Quebec, Canada
| | - Shirin A Enger
- Medical Physics Unit, Department of Oncology, Faculty of Medicine, McGill University, Montreal, Quebec, Canada
- Lady Davis Institute for Medical Research, Jewish General Hospital, Montreal, Quebec, Canada
| |
Collapse
|
2
|
Feasibility of electronic brachytherapy in cervix cancer-A dosimetric comparison of different brachytherapy techniques. Brachytherapy 2022; 21:389-396. [PMID: 35246391 DOI: 10.1016/j.brachy.2022.01.006] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 10/20/2021] [Revised: 01/03/2022] [Accepted: 01/04/2022] [Indexed: 11/23/2022]
Abstract
INTRODUCTION This study analyzes cases in which electronic brachytherapy (eBT) led to acceptable treatment plans in cervical cancer. Findings were compared with dosimetry values obtained in 192Ir-based treatments according to the high-risk clinical target volume (HR-CTV) and the disease stage. MATERIAL AND METHODS We retrospectively analyzed 48 patients with cervical cancer from two centers. The patients were treated with 192Ir based on MRI. It was possible to use interstitial needles via an Utrecht-type applicator. Dosimetry was simulated using eBT and the parameters D90 and D98 (HR-CTV) and D2cc, D1cc, and D0.1cc (bladder, rectum, and sigmoid colon) were evaluated. The Mann-Whitney U test was used for comparison. The overall cohort of patients was analyzed, as were the sub-cohorts based on stage (FIGO stages I+IIA, IIB and III-IV). Finally, the dosimetry of the eBT plans was evaluated, and the plans obtained were classified as "good", "acceptable", or "poor". RESULTS Statistically significant differences were found between the eBT and 192Ir plans for D98 (HR-CTV), D1cc and D0.1cc (bladder), and D1cc and D0.1cc (sigmoid colon). A total of 31 cases (64.6%) were considered good, seven (14.6%) were considered acceptable, and 10 (20.8%) were considered poor. For volumes <30 cc, all the plans were good or acceptable; for volumes >30 cc, 54.3% were good, and 71.4% were good or acceptable. By stage, eBT plans for patients with stage IB-IIA disease were good in 100%, whereas those for patients with stage IIB were good in 70.6% and III-IV disease were good in 50%. CONCLUSIONS eBT provides appropriate dosimetry for treatment of cervical cancer in selected cases.
Collapse
|
3
|
Ayala Alvarez DS, Watson PGF, Popovic M, Heng VJ, Evans MDC, Seuntjens J. Monte Carlo calculation of the TG-43 dosimetry parameters for the INTRABEAM source with spherical applicators. Phys Med Biol 2021; 66. [PMID: 34663769 DOI: 10.1088/1361-6560/ac309f] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 07/23/2021] [Accepted: 10/18/2021] [Indexed: 11/12/2022]
Abstract
OBJECTIVE The relative TG-43 dosimetry parameters of the INTRABEAM (Carl Zeiss Meditec AG, Jena, Germany) bare probe were recently reported by Ayala Alvarezet al(2020Phys. Med. Biol.65245041). The current study focuses on the dosimetry characterization of the INTRABEAM source with the eight available spherical applicators according to the TG-43 formalism using Monte Carlo (MC) simulations. APPROACH This report includes the calculated dose-rate conversion coefficients that determine the absolute dose rate to water at a reference point of 10 mm from the applicator surface, based on calibration air-kerma rate measurements at 50 cm from the source on its transverse plane. Since the air-kerma rate measurements are not yet provided from a standards laboratory for the INTRABEAM, the values in the present study were calculated with MC. This approach is aligned with other works in the search for standardization of the dosimetry of electronic brachytherapy sources. As a validation of the MC model, depth dose calculations along the source axis were compared with calibration data from the source manufacturer. MAIN RESULTS The calculated dose-rate conversion coefficients were 434.0 for the bare probe, and 683.5, 548.3, 449.9, 376.5, 251.0, 225.6, 202.8, and 182.6 for the source with applicators of increasing diameter from 15 to 50 mm, respectively. The radial dose and the 2D anisotropy functions of the TG-43 formalism were also obtained and tabulated in this document. SIGNIFICANCE This work presents the data required by a treatment planning system for the characterization of the INTRABEAM system in the context of intraoperative radiotherapy applications.
Collapse
Affiliation(s)
| | - Peter G F Watson
- Medical Physics Unit, McGill University and Cedars Cancer Center, Montreal, Canada
| | - Marija Popovic
- Medical Physics Unit, McGill University and Cedars Cancer Center, Montreal, Canada
| | - Veng Jean Heng
- Medical Physics Unit, McGill University and Cedars Cancer Center, Montreal, Canada
| | - Michael D C Evans
- Medical Physics Unit, McGill University and Cedars Cancer Center, Montreal, Canada
| | - Jan Seuntjens
- Medical Physics Unit, McGill University and Cedars Cancer Center, Montreal, Canada
| |
Collapse
|
4
|
Badali DS, Vainer Y, Ellenor CW, Mitchell CR, Fishman K, Soro N, Price R, Funk T. Inverse treatment planning for an electronic brachytherapy system delivering anisotropic radiation therapy. Phys Med Biol 2021; 66:055004. [PMID: 33429370 DOI: 10.1088/1361-6560/abda9a] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/11/2022]
Abstract
An inverse radiation treatment planning algorithm for Sensus Healthcare's SculpturaTM electronic brachytherapy system has been designed. The algorithm makes use of simulated annealing to optimize the conformation number (CN) of the treatment plan. The highly anisotropic dose distributions produced by the SculpturaTM x-ray source empower the inverse treatment planning algorithm to achieve highly conformal treatment plans for a wide range of prescribed planning target volumes. Over a set of 10 datasets the algorithm achieved an average CN of 0.79 ± 0.08 and an average gamma passing rate of 0.90 ± 0.10 at 5%/5 mm. A regularization term that encouraged short treatment plans was used, and it was found that the total treatment time could be reduced by 20% with only a nominal reduction in the CN and gamma passing rate. It was also found that downsampling the voxelized volume (from 3203 to 643 voxels) prior to optimization resulted in a 150× speedup in the optimization time (from 2 + minutes to < 1 s) without affecting the quality of the treatment plan.
Collapse
|
5
|
Lozares S, Font JA, Gandía A, Campos A, Flamarique S, Ibáñez R, Villa D, Alba V, Jiménez S, Hernández M, Casamayor C, Vicente I, Hernando E, Rubio P. In vivo dosimetry in low-voltage IORT breast treatments with XR-RV3 radiochromic film. Phys Med 2021; 81:173-181. [PMID: 33465753 DOI: 10.1016/j.ejmp.2020.12.011] [Citation(s) in RCA: 6] [Impact Index Per Article: 2.0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 08/21/2020] [Revised: 12/12/2020] [Accepted: 12/16/2020] [Indexed: 12/30/2022] Open
Abstract
PURPOSE The objectives of the study were to establish a procedure for in vivo film-based dosimetry for intraoperative radiotherapy (IORT), evaluate the typical doses delivered to organs at risk, and verify the dose prescription. MATERIALS AND METHODS In vivo dose measurements were studied using XR-RV3 radiochromic films in 30 patients with breast cancer undergoing IORT using the Axxent® device (Xoft Inc.). The stability of the radiochromic films in the energy ranges used was verified by taking measurements at different depths. The stability of the scanner response was tested, and 5 different calibration curves were constructed for different beam qualities. Six pieces of film were placed in each of the 30 patients. All the pieces were correctly sterilized and checked to ensure that the process did not affect the outcome. All calibration and dose measurements were analyzed using the Radiochromic.com software application. RESULTS The doses were measured for 30 patients. The doses in contact with the applicator (prescription zone) were 19.8 ± 0.9 Gy. In the skin areas, the doses were as follows: 1-2 cm from the applicator, 1.86 ± 0.77 Gy; 2-5 cm, 0.73 ± 0.14 Gy; and greater than 5 cm, 0.28 ± 0.17 Gy. The dose delivered to the pectoral muscle (tungsten shielding disc) was 0.51 ± 0.27 Gy. CONCLUSIONS The study demonstrated the viability of XR-RV3 films for in vivo dose measurement in the dose and energy ranges applied in a complex procedure, such as breast IORT. The doses in organs at risk were far below the tolerances for cases such as those studied.
Collapse
Affiliation(s)
- Sergio Lozares
- Medical Physics Department. Miguel Servet University Hospital Zaragoza, Spain.
| | - Jose A Font
- Medical Physics Department. Miguel Servet University Hospital Zaragoza, Spain
| | - Almudena Gandía
- Medical Physics Department. Miguel Servet University Hospital Zaragoza, Spain
| | - Arantxa Campos
- Radiation Oncology Department. Miguel Servet University Hospital Zaragoza, Spain
| | - Sonia Flamarique
- Radiation Oncology Department. Miguel Servet University Hospital Zaragoza, Spain
| | - Reyes Ibáñez
- Radiation Oncology Department. Miguel Servet University Hospital Zaragoza, Spain
| | - David Villa
- Medical Physics Department. Miguel Servet University Hospital Zaragoza, Spain
| | - Verónica Alba
- Medical Physics Department. Miguel Servet University Hospital Zaragoza, Spain
| | - Sara Jiménez
- Medical Physics Department. Miguel Servet University Hospital Zaragoza, Spain
| | - Mónica Hernández
- Medical Physics Department. Miguel Servet University Hospital Zaragoza, Spain
| | - Carmen Casamayor
- Endocrine, Bariatric and Breast Surgery Unit. General and Digestive Surgery Department. Miguel Servet University Hospital Zaragoza, Spain
| | - Isabel Vicente
- Breast Unit. Gynaecology Department. Miguel Servet University Hospital Zaragoza, Spain
| | - Ernesto Hernando
- Endocrine, Bariatric and Breast Surgery Unit. General and Digestive Surgery Department. Miguel Servet University Hospital Zaragoza, Spain
| | - Patricia Rubio
- Breast Unit. Gynaecology Department. Miguel Servet University Hospital Zaragoza, Spain
| |
Collapse
|
6
|
Ayala Alvarez DS, G F Watson P, Popovic M, Jean Heng V, Evans MDC, Seuntjens J. Monte Carlo calculation of the relative TG-43 dosimetry parameters for the INTRABEAM electronic brachytherapy source. Phys Med Biol 2020; 65:245041. [PMID: 33137796 DOI: 10.1088/1361-6560/abc6f1] [Citation(s) in RCA: 2] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/11/2022]
Abstract
The INTRABEAM system (Carl Zeiss Meditec AG, Jena, Germany) is an electronic brachytherapy (eBT) device designed for intraoperative radiotherapy applications. To date, the INTRABEAM x-ray source has not been characterized according to the AAPM TG-43 specifications for brachytherapy sources. This restricts its modelling in commercial treatment planning systems (TPSs), with the consequence that the doses to organs at risk are unknown. The aim of this work is to characterize the INTRABEAM source according to the TG-43 brachytherapy dosimetry protocol. The dose distribution in water around the source was determined with Monte Carlo (MC) calculations. For the validation of the MC model, depth dose calculations along the source longitudinal axis were compared with measurements using a soft x-ray ionization chamber (PTW 34013) and two synthetic diamond detectors (microDiamond PTW TN60019). In our results, the measurements in water agreed with the MC model calculations within uncertainties. The use of the microDiamond detector yielded better agreement with MC calculations, within estimated uncertainties, compared to the ionization chamber at points of steeper dose gradients. The radial dose function showed a steep fall-off close to the INTRABEAM source ([Formula: see text]10 mm) with a gradient higher than that of commonly used brachytherapy radionuclides (192Ir, 125I and 103Pd), with values of 2.510, 1.645 and 1.232 at 4, 6 and 8 mm, respectively. The radial dose function partially flattens at larger distances with a fall-off comparable to that of the Xoft Axxent® (iCAD, Inc., Nashua, NH) eBT system. The simulated 2D polar anisotropy close to the bare probe walls showed deviations from unity of up to 55% at 10 mm and 155°. This work presents the MC calculated TG-43 parameters for the INTRABEAM, which constitute the necessary data for the characterization of the source as required by a TPS used in clinical dose calculations.
Collapse
Affiliation(s)
| | - Peter G F Watson
- Medical Physics Unit, McGill University and Cedars Cancer Center, Montreal, Canada
| | - Marija Popovic
- Medical Physics Unit, McGill University and Cedars Cancer Center, Montreal, Canada
| | - Veng Jean Heng
- Medical Physics Unit, McGill University and Cedars Cancer Center, Montreal, Canada
| | - Michael D C Evans
- Medical Physics Unit, McGill University and Cedars Cancer Center, Montreal, Canada
| | - Jan Seuntjens
- Medical Physics Unit, McGill University and Cedars Cancer Center, Montreal, Canada
| |
Collapse
|
7
|
Watson PGF, Popovic M, Liang L, Tomic N, Devic S, Seuntjens J. Clinical Implication of Dosimetry Formalisms for Electronic Low-Energy Photon Intraoperative Radiation Therapy. Pract Radiat Oncol 2020; 11:e114-e121. [PMID: 32795615 DOI: 10.1016/j.prro.2020.07.005] [Citation(s) in RCA: 7] [Impact Index Per Article: 1.8] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 10/24/2019] [Revised: 05/27/2020] [Accepted: 07/27/2020] [Indexed: 11/26/2022]
Abstract
PURPOSE Intraoperative radiation therapy (IORT) using the INTRABEAM, a miniature x-ray source, has shown to be effective in treating breast cancer. However, recent investigations have suggested a significant deviation between the reported and delivered doses. In this work, the dose delivered by INTRABEAM in the TARGIT breast protocol was investigated, along with the dose from the Xoft Axxent, another source used in breast IORT. METHODS AND MATERIALS The absorbed dose from the INTRABEAM was determined from ionization chamber measurements using: (a) the manufacturer-recommended formula (Zeiss V4.0 method), (b) a Monte Carlo calculated chamber conversion factor (CQ method), and (c) the formula consistent with the TARGIT breast protocol (TARGIT method). The dose from the Xoft Axxent was determined from ionization chamber measurements using the Zeiss V4.0 method and calculated using the American Association of Physicists in Medicine TG-43 formalism. RESULTS For a nominal TARGIT prescription of 20 Gy, the dose at the INTRABEAM applicator surface ranged from 25.2 to 31.7 Gy according to the CQ method for the largest (5 cm) and smallest (1.5 cm) diameter applicator, respectively. The Zeiss V4.0 method results were 7% to 10% lower (23.2 to 28.6 Gy). At 1 cm depth, the CQ and Zeiss V4.0 absorbed doses were also larger than those predicted by the TARGIT method. The dose at 1 cm depth from the Xoft Axxent for a surface dose of 20 Gy was slightly less than INTRABEAM (3%-7% compared with CQ method). An exception was for the 3 cm applicator, where the Xoft dose was appreciably lower (31%). CONCLUSIONS The doses delivered in the TARGIT breast protocol with INTRABEAM were significantly greater than the prescribed 20 Gy and depended on the size of spherical applicator used. Breast IORT treatments with the Xoft Axxent received less dose compared with TARGIT INTRABEAM, which could have implications for studies comparing clinical outcomes between the 2 devices.
Collapse
Affiliation(s)
| | - Marija Popovic
- Medical Physics Unit, McGill University, Montreal, QC, Canada
| | - Liheng Liang
- Medical Physics Unit, Department of Radiation Oncology, SMBD Jewish General Hospital, McGill University, Montreal, QC, Canada
| | - Nada Tomic
- Medical Physics Unit, Department of Radiation Oncology, SMBD Jewish General Hospital, McGill University, Montreal, QC, Canada
| | - Slobodan Devic
- Medical Physics Unit, Department of Radiation Oncology, SMBD Jewish General Hospital, McGill University, Montreal, QC, Canada
| | - Jan Seuntjens
- Medical Physics Unit, McGill University, Montreal, QC, Canada
| |
Collapse
|
8
|
Shaikh MY, Tanny S. Dosimetric comparison of the INTRABEAM and Axxent for intraoperative breast radiotherapy. Brachytherapy 2020; 19:234-240. [DOI: 10.1016/j.brachy.2019.11.005] [Citation(s) in RCA: 4] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 06/06/2019] [Revised: 11/18/2019] [Accepted: 11/19/2019] [Indexed: 01/28/2023]
|
9
|
Martin E, Sowards K, Wang B. Investigation of a source model for a new electronic brachytherapy tandem by film measurement. J Appl Clin Med Phys 2018; 19:640-650. [PMID: 30105891 PMCID: PMC6123150 DOI: 10.1002/acm2.12440] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.2] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 04/23/2018] [Revised: 06/01/2018] [Accepted: 07/16/2018] [Indexed: 12/26/2022] Open
Abstract
PURPOSE To investigate the accuracy of a vendor-supplied source model for a new Xoft Axxent 0-degree titanium tandem by film measurement. METHODS We measured the anisotropy factors at varying distances and angles from the tandem in water using radiochromic film (Gafchromic EBT3) and an Epson Perfection v750 desktop flatbed scanner (US Epson, Long Beach, CA). A 0-degree tandem was placed vertically in a water phantom. Four pieces of film, each at varying depths, were positioned orthogonal to the longitudinal axis of the tandem for azimuthal anisotropy measurements. Polar anisotropy measurements were taken with the film aligned parallel to the tandem. An absolute dose calibration for the film was verified with a PTW 34013 Soft X-Ray Chamber. The film measurements were analyzed using different color channels. The measured polar anisotropy for varying source positions was compared to the vendor's data. Azimuthal anisotropy was measured as a function of the radius and angle, and normalized to the mean value over all angles at the specified radius. RESULTS The azimuthal anisotropy of the tandem and source was found to be consistent for different positions along the tandem's longitudinal axis and at varying distances from the tandem. Absolute dose using a calibrated parallel plate chamber showed agreement to within 2% of expected TPS values. The custom tandem, which has a thicker tip than the wall, was attenuating the 50 kV photons more than expected, at the angles where the photons had more wall material to traverse. This discrepancy was verified at different distances from the tandem and with different measurement techniques. As distance increased, anisotropy values had better agreement. CONCLUSIONS We quantified the agreement between the measured and provided anisotropy factors for a new Xoft Axxent 0-degree titanium tandem. Radiochromic film response at low kV energy was also investigated. Our results showed that vendor-supplied TG-43 values were appropriate for clinical use at majority of the angles. A rigorous quality assurance method for new electronic brachytherapy sources and applicators, along with complete knowledge of all dosimetric measuring tools, should be implemented for all parts of the verification and commissioning process.
Collapse
Affiliation(s)
- Elijah Martin
- Department of Radiation OncologyUniversity of LouisvilleLouisvilleKYUSA
| | - Keith Sowards
- Department of Radiation OncologyUniversity of LouisvilleLouisvilleKYUSA
| | - Brian Wang
- Department of Radiation OncologyUniversity of LouisvilleLouisvilleKYUSA
| |
Collapse
|
10
|
Barati B, Zabihzadeh M, Tahmasebi Birgani M, Chegini N, Fatahiasl J, Mirr I. Evaluation of the Effect of Source Geometry on the Output of Miniature X-ray Tube for Electronic Brachytherapy through Simulation. J Biomed Phys Eng 2018; 8:29-42. [PMID: 29732338 PMCID: PMC5928308] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 12/05/2016] [Accepted: 01/21/2017] [Indexed: 11/08/2022]
Abstract
OBJECTIVE The use of miniature X-ray source in electronic brachytherapy is on the rise so there is an urgent need to acquire more knowledge on X-ray spectrum production and distribution by a dose. The aim of this research was to investigate the influence of target thickness and geometry at the source of miniature X-ray tube on tube output. METHOD Five sources were simulated based on problems each with a specific geometric structure and conditions using MCNPX code. Tallies proportional to the output were used to calculate the results for the influence of source geometry on output. RESULTS The results of this work include the size of the optimal thickness of 5 miniature sources, energy spectrum of the sources per 50 kev and also the axial and transverse dose of simulated sources were calculated based on these thicknesses. The miniature source geometric was affected on the output x-ray tube. CONCLUSION The result of this study demonstrates that hemispherical-conical, hemispherical and truncated-conical miniature sources were determined as the most suitable tools.
Collapse
Affiliation(s)
- B. Barati
- Department of Medical Physics, School of Medicine, Ahvaz Jundishapur University of Medical Sciences, Ahvaz, Iran
| | - M. Zabihzadeh
- Department of Medical Physics, School of Medicine, Ahvaz Jundishapur University of Medical Sciences, Ahvaz, Iran
| | - M.J. Tahmasebi Birgani
- Department of Medical Physics, School of Medicine, Ahvaz Jundishapur University of Medical Sciences, Ahvaz, Iran
,Department of Clinical Oncology, Golestan Hospital, Ahvaz Jundishapur University of Medical Sciences, Ahvaz, Iran
| | - N. Chegini
- Department of Medical Physics, School of Medicine, Ahvaz Jundishapur University of Medical Sciences, Ahvaz, Iran
| | - J. Fatahiasl
- Department of Medical Physics, School of Medicine, Ahvaz Jundishapur University of Medical Sciences, Ahvaz, Iran
| | - I. Mirr
- Department of Biostatistics and Epidemiology, School of Health, Ahvaz Jundishapur University of Medical Sciences, Ahvaz, Iran
| |
Collapse
|
11
|
Redler G, Templeton A, Zhen H, Turian J, Bernard D, Chu JC, Griem KL, Liao Y. Dosimetric effects of saline- versus water-filled balloon applicators for IORT using the model S700 electronic brachytherapy source. Brachytherapy 2018; 17:500-505. [DOI: 10.1016/j.brachy.2017.11.003] [Citation(s) in RCA: 4] [Impact Index Per Article: 0.7] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 09/12/2017] [Revised: 11/03/2017] [Accepted: 11/03/2017] [Indexed: 12/19/2022]
|
12
|
Barati B, Zabihzadeh M, Birgani MJT, Chegini N, Ghahfarokhi MH, Fatahiasl J. Assessment of two hemispherical and hemispherical-conical miniature sources used in electronic brachytherapy using Monte Carlo Simulation. Electron Physician 2017; 9:3845-3856. [PMID: 28465817 PMCID: PMC5410916 DOI: 10.19082/3845] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.1] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 09/27/2016] [Accepted: 12/22/2016] [Indexed: 01/29/2023] Open
Abstract
Introduction Since the heart of the electronic brachytherapy system is a tube of a miniature x-ray and due to the increasing use of electronic brachytherapy, there is an urgent need for acquiring knowledge about the X-ray spectrum produced, and distribution of x-ray dose. This study aimed to assess the optimal target thickness (TT), the X-ray source spectrum, and the absorbed dose of two miniature sources of hemispherical and hemispherical-conical used in electronic brachytherapy systems, through a Monte Carlo simulation. Methods Considering the advantages of MCNPX Code (2.6.0), two input files corresponding to the characteristics of the investigated miniature sources were prepared for this code and then were used for simulation. The optimal thickness (OT) of gold and tungsten targets was determined for the energy levels of 40, 45, and 50 kilo-electron-volts. Results In this study, the values of the size of the optimal thickness of 0.92, 1.01 and 1.06 μ for gold target and values of 0.99, 1.08 and 1.34 μ for tungsten target were obtained for energies 40, 45 and 50 keV that using these values, the optimum thickness of 0.92, X-ray spectrum within and outside targets, axial and radial doses for the used energy were calculated for two miniature sources. Conclusion It was found that the energy of incident electron, target shape, cross-sectional area of the produced bremsstrahlung, atomic number of materials constituting of the target and output window are the factors with the greatest impacts on the produced X-ray spectrum and the absorbed dose.
Collapse
Affiliation(s)
- Barat Barati
- Ph.D. Candidate, Department of Medical Physics, School of Medicine, Ahvaz Jundishapur University of Medical Sciences, Ahvaz, Iran
| | - Mansour Zabihzadeh
- Ph.D. of Medical Physics, Assistant Professor, Department of Medical Physics, School of Medicine, Ahvaz Jundishapur University of Medical Sciences, Ahvaz, Iran
| | - Mohammad Javad Tahmasebi Birgani
- Ph.D. of Medical Physics, Professor, Department of Medical Physics, School of Medicine, Ahvaz Jundishapur University of Medical Sciences, Ahvaz, Iran
| | - Nahid Chegini
- Ph.D. of Medical Physics, Assistant Professor, Department of Medical Physics, School of Medicine, Ahvaz Jundishapur University of Medical Sciences, Ahvaz, Iran
| | - Mojtaba Hoseini Ghahfarokhi
- Ph.D. Candidate, Member of Student Research Committee, Department of Medical Physics, School of Medicine, Ahvaz Jundishapur University of Medical Sciences, Ahvaz, Iran
| | - Jafar Fatahiasl
- Ph.D. of Medical Physics, Assistant Professor, Department of Para Medicine, School of Medicine, Ahvaz Jundishapur University of Medical Sciences, Ahvaz, Iran
| |
Collapse
|
13
|
Hiatt JR, Rivard MJ, Hughes HG. Simulation evaluation of NIST air‐kerma rate calibration standard for electronic brachytherapy. Med Phys 2016; 43:1119-29. [DOI: 10.1118/1.4940791] [Citation(s) in RCA: 8] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/26/2022] Open
Affiliation(s)
- Jessica R. Hiatt
- Biomedical Engineering and Biotechnology, University Massachusetts Lowell, Lowell, Massachusetts 01854
| | - Mark J. Rivard
- Department of Radiation Oncology, Tufts University School of Medicine, Boston, Massachusetts 02111
| | - H. Grady Hughes
- Los Alamos National Laboratory, Los Alamos, New Mexico 87545
| |
Collapse
|
14
|
Hiatt JR, Davis SD, Rivard MJ. A revised dosimetric characterization of the model S700 electronic brachytherapy source containing an anode-centering plastic insert and other components not included in the 2006 model. Med Phys 2015; 42:2764-76. [DOI: 10.1118/1.4919280] [Citation(s) in RCA: 12] [Impact Index Per Article: 1.3] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/15/2022] Open
|