Archibald-Heeren B, Liu G. Raystation Monte Carlo application: evaluation of electron calculations with entry obliquity.
AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE 2016;
39:441-52. [PMID:
27052438 DOI:
10.1007/s13246-016-0437-y]
[Citation(s) in RCA: 6] [Impact Index Per Article: 0.8] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Received: 08/18/2015] [Accepted: 03/10/2016] [Indexed: 11/26/2022]
Abstract
To evaluate the accuracy of Raystation's implementation for Monte Carlo VMC ++ with electrons at varying angles of incidence for low and medium energy electron beams. Thirty-two profile and percentage depth dose scans were taken at 5° incident angle intervals for 6 and 12 MeV and compared to extracted fluences from Raystation calculations using gamma analysis with 2 %/2 mm criteria. Point dose measurements were compared to calculated doses to determine output accuracy. Electron profile and percentage depth dose curves for both energies show good agreement between 0° and 20° with 29/30 scans above 90 % pass rate. Average variation between calculated and measured point doses was -0.73 % with all measurements falling within ±2 % of calculated dose. Raystation's application of VMC ++ Monte Carlo algorithm provides clinically acceptable accuracy for low and medium energy electron dosimetry at incident angles up to 20° for Varian Clinac iX models.