1
|
Werner R, Szkitsak J, Sentker T, Madesta F, Schwarz A, Fernolendt S, Vornehm M, Gauer T, Bert C, Hofmann C. Comparison of intelligent 4D CT sequence scanning and conventional spiral 4D CT: a first comprehensive phantom study. Phys Med Biol 2021; 66. [PMID: 33171441 DOI: 10.1088/1361-6560/abc93a] [Citation(s) in RCA: 8] [Impact Index Per Article: 2.7] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 08/05/2020] [Accepted: 11/10/2020] [Indexed: 11/11/2022]
Abstract
4D CT imaging is a cornerstone of 4D radiotherapy treatment. Clinical 4D CT data are, however, often affected by severe artifacts. The artifacts are mainly caused by breathing irregularity and retrospective correlation of breathing phase information and acquired projection data, which leads to insufficient projection data coverage to allow for proper reconstruction of 4D CT phase images. The recently introduced 4D CT approach i4DCT (intelligent 4D CT sequence scanning) aims to overcome this problem by breathing signal-driven tube control. The present motion phantom study describes the first in-depth evaluation of i4DCT in a real-world scenario. Twenty-eight 4D CT breathing curves of lung and liver tumor patients with pronounced breathing irregularity were selected to program the motion phantom. For every motion pattern, 4D CT imaging was performed with i4DCT and a conventional spiral 4D CT mode. For qualitative evaluation, the reconstructed 4D CT images were presented to clinical experts, who scored image quality. Further quantitative evaluation was based on established image intensity-based artifact metrics to measure (dis)similarity of neighboring image slices. In addition, beam-on and scan times of the scan modes were analyzed. The expert rating revealed a significantly higher image quality for the i4DCT data. The quantitative evaluation further supported the qualitative: While 20% of the slices of the conventional spiral 4D CT images were found to be artifact-affected, the corresponding fraction was only 4% for i4DCT. The beam-on time (surrogate of imaging dose) did not significantly differ between i4DCT and spiral 4D CT. Overall i4DCT scan times (time between first beam-on and last beam-on event, including scan breaks to compensate for breathing irregularity) were, on average, 53% longer compared to spiral CT. Thus, the results underline that i4DCT significantly improves 4D CT image quality compared to standard spiral CT scanning in the case of breathing irregularity during scanning.
Collapse
Affiliation(s)
- René Werner
- University Medical Center Hamburg-Eppendorf, Martinistr. 52, 20246 Hamburg, Germany
| | - Juliane Szkitsak
- Department of Radiation Oncology, Universitätsklinikum Erlangen, Friederich-Alexander-Universität Erlangen-Nürnberg, 91504 Erlangen, Germany
| | - Thilo Sentker
- University Medical Center Hamburg-Eppendorf, Martinistr. 52, 20246 Hamburg, Germany
| | - Frederic Madesta
- University Medical Center Hamburg-Eppendorf, Martinistr. 52, 20246 Hamburg, Germany
| | - Annette Schwarz
- Friederich-Alexander-Universität Erlangen-Nürnberg, 91504 Erlangen, Germany.,Siemens Healthcare GmbH, Siemensstr. 3, 91301 Forchheim, Germany
| | - Susanne Fernolendt
- Friederich-Alexander-Universität Erlangen-Nürnberg, 91504 Erlangen, Germany.,Siemens Healthcare GmbH, Siemensstr. 3, 91301 Forchheim, Germany
| | - Marc Vornehm
- Friederich-Alexander-Universität Erlangen-Nürnberg, 91504 Erlangen, Germany.,Siemens Healthcare GmbH, Siemensstr. 3, 91301 Forchheim, Germany
| | - Tobias Gauer
- University Medical Center Hamburg-Eppendorf, Martinistr. 52, 20246 Hamburg, Germany
| | - Christoph Bert
- Department of Radiation Oncology, Universitätsklinikum Erlangen, Friederich-Alexander-Universität Erlangen-Nürnberg, 91504 Erlangen, Germany
| | | |
Collapse
|
2
|
Du Q, Baine M, Bavitz K, McAllister J, Liang X, Yu H, Ryckman J, Yu L, Jiang H, Zhou S, Zhang C, Zheng D. Radiomic feature stability across 4D respiratory phases and its impact on lung tumor prognosis prediction. PLoS One 2019; 14:e0216480. [PMID: 31063500 PMCID: PMC6504105 DOI: 10.1371/journal.pone.0216480] [Citation(s) in RCA: 28] [Impact Index Per Article: 5.6] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 11/12/2018] [Accepted: 04/22/2019] [Indexed: 11/25/2022] Open
Abstract
Radiomic analysis has recently demonstrated versatile uses in improving diagnostic and prognostic prediction accuracy for lung cancer. However, since lung tumors are subject to substantial motion due to respiration, the stability of radiomic features over the respiratory cycle of the patient needs to be investigated to better evaluate the robustness of the inter-patient feature variability for clinical applications, and its impact in such applications needs to be assessed. A full panel of 841 radiomic features, including tumor intensity, shape, texture, and wavelet features, were extracted from individual phases of a four-dimensional (4D) computed tomography on 20 early-stage non-small-cell lung cancer (NSCLC) patients. The stability of each radiomic feature was assessed across different phase images of the same patient using the coefficient of variation (COV). The relationship between individual COVs and tumor motion magnitude was inspected. Population COVs, the mean COVs of all 20 patients, were used to evaluate feature motion stability and categorize the radiomic features into 4 different groups. The two extremes, the Very Small group (COV≤5%) and the Large group (COV>20%), each accounted for about a quarter of the features. Shape features were the most stable, with COV≤10% for all features. A clinical study was subsequently conducted using 140 early-stage NSCLC patients. Radiomic features were employed to predict the overall survival with a 500-round bootstrapping. Identical multiple regression model development process was applied, and the model performance was compared between models with and without a feature pre-selection step based on 4D COV to pre-exclude unstable features. Among the systematically tested cutoff values, feature pre-selection with 4D COV≤5% achieved the optimal model performance. The resulting 3-feature radiomic model significantly outperformed its counterpart with no 4D COV pre-selection, with P = 2.16x10-27 in the one-tailed t-test comparing the prediction performances of the two models.
Collapse
Affiliation(s)
- Qian Du
- Biological Sciences, University of Nebraska Lincoln, Lincoln, NE, United States of America
| | - Michael Baine
- Radiation Oncology, University of Nebraska Medical Center, Omaha, NE, United States of America
| | - Kyle Bavitz
- Radiation Oncology, University of Nebraska Medical Center, Omaha, NE, United States of America
| | - Josiah McAllister
- Radiation Oncology, University of Nebraska Medical Center, Omaha, NE, United States of America
| | - Xiaoying Liang
- University of Florida Proton Therapy Institute, Jacksonville, FL, United States of America
| | - Hongfeng Yu
- Computer Science, University of Nebraska Lincoln, Lincoln, NE, United States of America
| | - Jeffrey Ryckman
- Radiation Oncology, University of Nebraska Medical Center, Omaha, NE, United States of America
| | - Lina Yu
- Computer Science, University of Nebraska Lincoln, Lincoln, NE, United States of America
| | - Hengle Jiang
- Computer Science, University of Nebraska Lincoln, Lincoln, NE, United States of America
| | - Sumin Zhou
- Radiation Oncology, University of Nebraska Medical Center, Omaha, NE, United States of America
| | - Chi Zhang
- Biological Sciences, University of Nebraska Lincoln, Lincoln, NE, United States of America
| | - Dandan Zheng
- Radiation Oncology, University of Nebraska Medical Center, Omaha, NE, United States of America
- * E-mail:
| |
Collapse
|
3
|
Tanaka S, Kadoya N, Kajikawa T, Matsuda S, Dobashi S, Takeda K, Jingu K. Investigation of thoracic four-dimensional CT-based dimension reduction technique for extracting the robust radiomic features. Phys Med 2019; 58:141-148. [PMID: 30824145 DOI: 10.1016/j.ejmp.2019.02.009] [Citation(s) in RCA: 12] [Impact Index Per Article: 2.4] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 09/14/2018] [Revised: 01/29/2019] [Accepted: 02/12/2019] [Indexed: 11/25/2022] Open
Abstract
Robust feature selection in radiomic analysis is often implemented using the RIDER test-retest datasets. However, the CT Protocol between the facility and test-retest datasets are different. Therefore, we investigated possibility to select robust features using thoracic four-dimensional CT (4D-CT) scans that are available from patients receiving radiation therapy. In 4D-CT datasets of 14 lung cancer patients who underwent stereotactic body radiotherapy (SBRT) and 14 test-retest datasets of non-small cell lung cancer (NSCLC), 1170 radiomic features (shape: n = 16, statistics: n = 32, texture: n = 1122) were extracted. A concordance correlation coefficient (CCC) > 0.85 was used to select robust features. We compared the robust features in various 4D-CT group with those in test-retest. The total number of robust features was a range between 846/1170 (72%) and 970/1170 (83%) in all 4D-CT groups with three breathing phases (40%-60%); however, that was a range between 44/1170 (4%) and 476/1170 (41%) in all 4D-CT groups with 10 breathing phases. In test-retest, the total number of robust features was 967/1170 (83%); thus, the number of robust features in 4D-CT was almost equal to that in test-retest by using 40-60% breathing phases. In 4D-CT, respiratory motion is a factor that greatly affects the robustness of features, thus by using only 40-60% breathing phases, excessive dimension reduction will be able to be prevented in any 4D-CT datasets, and select robust features suitable for CT protocol of your own facility.
Collapse
Affiliation(s)
- Shohei Tanaka
- Department of Radiation Oncology, Tohoku University Graduate School of Medicine, Sendai, Japan
| | - Noriyuki Kadoya
- Department of Radiation Oncology, Tohoku University Graduate School of Medicine, Sendai, Japan.
| | - Tomohiro Kajikawa
- Department of Radiation Oncology, Tohoku University Graduate School of Medicine, Sendai, Japan
| | - Shohei Matsuda
- Department of Radiation Oncology, Tohoku University Graduate School of Medicine, Sendai, Japan
| | - Suguru Dobashi
- Department of Radiological Technology, School of Health Sciences, Faculty of Medicine, Tohoku University, Sendai, Japan
| | - Ken Takeda
- Department of Radiological Technology, School of Health Sciences, Faculty of Medicine, Tohoku University, Sendai, Japan
| | - Keiichi Jingu
- Department of Radiation Oncology, Tohoku University Graduate School of Medicine, Sendai, Japan
| |
Collapse
|
4
|
Liu G, Hu F, Ding X, Li X, Shao Q, Wang Y, Yang J, Quan H. Simulation of dosimetry impact of 4DCT uncertainty in 4D dose calculation for lung SBRT. Radiat Oncol 2019; 14:1. [PMID: 30621744 PMCID: PMC6323842 DOI: 10.1186/s13014-018-1191-y] [Citation(s) in RCA: 23] [Impact Index Per Article: 4.6] [Reference Citation Analysis] [Abstract] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 06/19/2018] [Accepted: 11/21/2018] [Indexed: 11/10/2022] Open
Abstract
Background Due to the heterogeneity of patient’s individual respiratory motion pattern in lung stereotactic body radiotherapy (SBRT), treatment planning dose assessment using a traditional four-dimensional computed tomography (4DCT_traditional) images based on a uniform breathing curve may not represent the true treatment dose delivered to the patient. The purpose of this study was to evaluate the accumulated dose discrepancy between based on the 4DCT_traditional and true 4DCT (4DCT_true) that incorporated with the patient’s real entire breathing motion. The study also explored a novel 4D robust planning strategy to compensate for such heterogeneity respiratory motion uncertainties. Methods Simulated and measured patient specific breathing curves were used to generate 4D targets motion CT images. Volumetric-modulated arc therapy (VMAT) was planned using two arcs. Accumulated dose was obtained by recalculating the plan dose on each individual phase image and then deformed the dose from each phase image to the reference image. The “4 D dose” (D4D) and “true dose” (Dtrue) were the accumulated dose based on the 4DCT_traditional and 4DCT_true respectively. The average worse case dose discrepancy (\documentclass[12pt]{minimal}
\usepackage{amsmath}
\usepackage{wasysym}
\usepackage{amsfonts}
\usepackage{amssymb}
\usepackage{amsbsy}
\usepackage{mathrsfs}
\usepackage{upgreek}
\setlength{\oddsidemargin}{-69pt}
\begin{document}$$ \overline{\Delta D} $$\end{document}ΔD¯) between D4D and Dtrue in all treatment fraction was calculated to evaluate dosimetric /planning parameters and correlate them with the heterogeneity of respiratory-induced motion patterns. A novel 4D robust optimization strategy for VMAT (4D Ro-VMAT) based on the probability density function(pdf) of breathing curve was proposed to improve the target coverage in the presence of heterogeneity respiratory motion. The data were assessed with a paired t-tests. Results With increasing breathing amplitude from 5 to 20 mm, target \documentclass[12pt]{minimal}
\usepackage{amsmath}
\usepackage{wasysym}
\usepackage{amsfonts}
\usepackage{amssymb}
\usepackage{amsbsy}
\usepackage{mathrsfs}
\usepackage{upgreek}
\setlength{\oddsidemargin}{-69pt}
\begin{document}$$ \overline{\Delta {D}_{99}} $$\end{document}ΔD99¯, \documentclass[12pt]{minimal}
\usepackage{amsmath}
\usepackage{wasysym}
\usepackage{amsfonts}
\usepackage{amssymb}
\usepackage{amsbsy}
\usepackage{mathrsfs}
\usepackage{upgreek}
\setlength{\oddsidemargin}{-69pt}
\begin{document}$$ \overline{\Delta {D}_{95}} $$\end{document}ΔD95¯ increased from 1.59,1.39 to 10.15%,8.66% respectively. When the standard deviation of breathing amplitude increased from 15 to 35% of the mean amplitude, \documentclass[12pt]{minimal}
\usepackage{amsmath}
\usepackage{wasysym}
\usepackage{amsfonts}
\usepackage{amssymb}
\usepackage{amsbsy}
\usepackage{mathrsfs}
\usepackage{upgreek}
\setlength{\oddsidemargin}{-69pt}
\begin{document}$$ \overline{\Delta {D}_{99}} $$\end{document}ΔD99¯, \documentclass[12pt]{minimal}
\usepackage{amsmath}
\usepackage{wasysym}
\usepackage{amsfonts}
\usepackage{amssymb}
\usepackage{amsbsy}
\usepackage{mathrsfs}
\usepackage{upgreek}
\setlength{\oddsidemargin}{-69pt}
\begin{document}$$ \overline{\Delta {D}_{95}} $$\end{document}ΔD95¯ increased from 4.06,3.48 to 10.25%,6.63% respectively. The 4D Ro-VMAT plan significantly improve the target dose compared to VMAT plan. Conclusion When the breathing curve amplitude is more than 10 mm and standard deviation of amplitude is higher than 25% of mean amplitude, special care is needed to choose an appropriated dose accumulation approach to evaluate lung SBRT plan target coverage robustness. The proposed 4D Ro_VMAT strategy based on the pdf of patient specific breathing curve could effectively compensate such uncertainties.
Collapse
Affiliation(s)
- Gang Liu
- Key Laboratory of Artificial Micro- and Nano- structures of Ministry of Education and Center for Electronic Microscopy, School of Physics and Technology, Wuhan University, Wuhan, 430072, China.,Cancer Center, Union Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan, 430022, China
| | - Fala Hu
- School of Mathematics and Statistics, Wuhan University, Wuhan, 430072, China
| | - Xuanfeng Ding
- Proton Therapy Center Beaumont Health, Royal Oak, MI, 48074, USA
| | - Xiaoqiang Li
- Proton Therapy Center Beaumont Health, Royal Oak, MI, 48074, USA
| | - Qihong Shao
- Wuhan Zhongyuan Electronics Group Co. LTD, Wuhan, 430205, China
| | - Yuenan Wang
- Cancer Hospital Chinese Academy of Medical Sciences, Shenzhen Center, Shenzhen, 518000, China
| | - Jing Yang
- Cancer Center, Union Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan, 430022, China
| | - Hong Quan
- Key Laboratory of Artificial Micro- and Nano- structures of Ministry of Education and Center for Electronic Microscopy, School of Physics and Technology, Wuhan University, Wuhan, 430072, China.
| |
Collapse
|
5
|
Hui C, Suh Y, Robertson D, Pan T, Das P, Crane CH, Beddar S. Internal respiratory surrogate in multislice 4D CT using a combination of Fourier transform and anatomical features. Med Phys 2015; 42:4338-48. [PMID: 26133631 DOI: 10.1118/1.4922692] [Citation(s) in RCA: 4] [Impact Index Per Article: 0.4] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/07/2022] Open
Abstract
PURPOSE The purpose of this study was to develop a novel algorithm to create a robust internal respiratory signal (IRS) for retrospective sorting of four-dimensional (4D) computed tomography (CT) images. METHODS The proposed algorithm combines information from the Fourier transform of the CT images and from internal anatomical features to form the IRS. The algorithm first extracts potential respiratory signals from low-frequency components in the Fourier space and selected anatomical features in the image space. A clustering algorithm then constructs groups of potential respiratory signals with similar temporal oscillation patterns. The clustered group with the largest number of similar signals is chosen to form the final IRS. To evaluate the performance of the proposed algorithm, the IRS was computed and compared with the external respiratory signal from the real-time position management (RPM) system on 80 patients. RESULTS In 72 (90%) of the 4D CT data sets tested, the IRS computed by the authors' proposed algorithm matched with the RPM signal based on their normalized cross correlation. For these data sets with matching respiratory signals, the average difference between the end inspiration times (Δtins) in the IRS and RPM signal was 0.11 s, and only 2.1% of Δtins were more than 0.5 s apart. In the eight (10%) 4D CT data sets in which the IRS and the RPM signal did not match, the average Δtins was 0.73 s in the nonmatching couch positions, and 35.4% of them had a Δtins greater than 0.5 s. At couch positions in which IRS did not match the RPM signal, a correlation-based metric indicated poorer matching of neighboring couch positions in the RPM-sorted images. This implied that, when IRS did not match the RPM signal, the images sorted using the IRS showed fewer artifacts than the clinical images sorted using the RPM signal. CONCLUSIONS The authors' proposed algorithm can generate robust IRSs that can be used for retrospective sorting of 4D CT data. The algorithm is completely automatic and requires very little processing time. The algorithm is cost efficient and can be easily adopted for everyday clinical use.
Collapse
Affiliation(s)
- Cheukkai Hui
- Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, Texas 77030
| | - Yelin Suh
- Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, Texas 77030
| | - Daniel Robertson
- Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, Texas 77030 and Department of Radiation Physics, The University of Texas Graduate School of Biomedical Sciences, Houston, Texas 77030
| | - Tinsu Pan
- Department of Imaging Physics, The University of Texas MD Anderson Cancer Center, Houston, Texas 77030 and Department of Imaging Physics, The University of Texas Graduate School of Biomedical Sciences, Houston, Texas 77030
| | - Prajnan Das
- Department of Radiation Oncology, The University of Texas MD Anderson Cancer Center, Houston, Texas 77030
| | - Christopher H Crane
- Department of Radiation Oncology, The University of Texas MD Anderson Cancer Center, Houston, Texas 77030
| | - Sam Beddar
- Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, Texas 77030 and Department of Radiation Physics, The University of Texas Graduate School of Biomedical Sciences, Houston, Texas 77030
| |
Collapse
|
6
|
Castillo SJ, Castillo R, Castillo E, Pan T, Ibbott G, Balter P, Hobbs B, Guerrero T. Evaluation of 4D CT acquisition methods designed to reduce artifacts. J Appl Clin Med Phys 2015; 16:4949. [PMID: 26103169 PMCID: PMC4504190 DOI: 10.1120/jacmp.v16i2.4949] [Citation(s) in RCA: 21] [Impact Index Per Article: 2.3] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 02/21/2014] [Revised: 11/21/2014] [Accepted: 11/09/2014] [Indexed: 12/25/2022] Open
Abstract
Four-dimensional computed tomography (4D CT) is used to account for respiratory motion in radiation treatment planning, but artifacts resulting from the acquisition and postprocessing limit its accuracy. We investigated the efficacy of three experimental 4D CT acquisition methods to reduce artifacts in a prospective institutional review board approved study. Eighteen thoracic patients scheduled to undergo radiation therapy received standard clinical 4D CT scans followed by each of the alternative 4D CT acquisitions: 1) data oversampling, 2) beam gating with breathing irregularities, and 3) rescanning the clinical acquisition acquired during irregular breathing. Relative values of a validated correlation-based artifact metric (CM) determined the best acquisition method per patient. Each 4D CT was processed by an extended phase sorting approach that optimizes the quantitative artifact metric (CM sorting). The clinical acquisitions were also postprocessed by phase sorting for artifact comparison of our current clinical implementation with the experimental methods. The oversampling acquisition achieved the lowest artifact presence among all acquisitions, achieving a 27% reduction from the current clinical 4D CT implementation (95% confidence interval = 34-20). The rescan method presented a significantly higher artifact presence from the clinical acquisition (37%; p < 0.002), the gating acquisition (26%; p < 0.005), and the oversampling acquisition (31%; p < 0.001), while the data lacked evidence of a significant difference between the clinical, gating, and oversampling methods. The oversampling acquisition reduced artifact presence from the current clinical 4D CT implementation to the largest degree and provided the simplest and most reproducible implementation. The rescan acquisition increased artifact presence significantly, compared to all acquisitions, and suffered from combination of data from independent scans over which large internal anatomic shifts occurred.
Collapse
|