McManus NH. Microtiter assay for interferon: microspectrophotometric quantitation of cytopathic effect.
Appl Environ Microbiol 1976;
31:35-8. [PMID:
182074 PMCID:
PMC169714 DOI:
10.1128/aem.31.1.35-38.1976]
[Citation(s) in RCA: 30] [Impact Index Per Article: 0.6] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/13/2022] Open
Abstract
This report describes an accurate, reproducible, and efficient microassay for human interferon, using a dye-uptake method to quantitate cytopathogenicity. The antiviral activity was measured by using a Gilford 300-N microsepctrophotometer, with the automatic programmer, sampler, and data lister. Two hundred interferon samples, each in a final volume of 1.0 ml, may be analyzed and recorded in 45 min. The reproducibility of a laboratory reference interferon and the human international reference B 69/19 on two different cell lines, using the model Q Oxford sampler, was found to be excellent, with the standard error of the log10 geometric mean of both references +/- 0.04 to 0.06.