Wexberg P, Kirisits C, Gyöngyösi M, Gottsauner-Wolf M, Ploner M, Pokrajac B, Pötter R, Glogar D. Vascular morphometric changes after radioactive stent implantation: a dose-response analysis.
J Am Coll Cardiol 2002;
39:400-7. [PMID:
11823076 DOI:
10.1016/s0735-1097(01)01776-4]
[Citation(s) in RCA: 2] [Impact Index Per Article: 0.1] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Indexed: 11/26/2022]
Abstract
OBJECTIVES
The goal of this study was to evaluate the dose-dependency of morphometric changes in the coronary arterial wall after radioactive stenting.
BACKGROUND
Radioactive stents have been found to reduce intrastent intimal hyperplasia (IIH) but lead to a characteristic type of restenosis occurring predominantly at the stent edges.
METHODS
Fifteen patients underwent intravascular ultrasound (IVUS) examination after implantation of a P-32 radioactive stent and at the six-month follow-up. The post-stent IVUS measurements on seven predefined locations of each lesion were subjected to a computer algorithm for the development of dose-volume histograms (DVH). Thus, we derived the radiation doses delivered to at least 10% and 90% of the adventitia (DV10, DV90). The IIH and vascular remodeling at follow-up were correlated with the doses in each segment.
RESULTS
The IIH was most pronounced at the stent edges and lowest in the stent-body, whereas we detected a significant expansive remodeling within the stent body. The delivered doses correlated with a decreased IIH (r = 0.52, p < 0.001 for DV10 and r = 0.62, p < 0.001 for DV90) and with expansive remodeling (r = 0.48, p = 0.009 for DV10 and r = 0.50, p = 0.006 for DV90). A DV10 >90 Gy or a DV90 >15 Gy reduced IIH and induced expansive remodeling. Plaque growth was not reduced by radioactive stents.
CONCLUSIONS
The DVH analysis reveals a dose-dependent increase of external elastic lamina area behind radioactive stents, whereas plaque growth is not reduced but inverted into an outward direction from the stent. A DV10 >90 Gy or a DV90 >15 Gy results in a beneficial long-term outcome after radioactive stenting.
Collapse