Bukovitz AG. Computer program for central axis dose calculations for high energy photons.
ACTA RADIOLOGICA: THERAPY, PHYSICS, BIOLOGY 1977;
16:427-32. [PMID:
414521 DOI:
10.3109/02841867709133964]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 12/15/2022]
Abstract
A computer program is described for central axis dose calculations for 60Co and 6 MV photons. Fixed distance or isocentric calculations for regular or irregular fields can be performed. Corrections are made for wedges and blocking trays. Tissue-air ratios and percentage depth doses are calculated from scatter-air ratios eliminating the need for empirical formulas. Calculated tissue-air ratios and percentage depth doses for elongated fields agree to within 2 per cent with measured data, eliminating the need for tables of equivalent squares. Calculated values for irregular fields also agree to within 2 per cent with measured data.