Bauer CAJ, Scheurer M, Bourauel C, Kretzer JP, Roser CJ, Lux CJ, Hodecker LD. Precision of slot widths and torque transmission of in-office 3D printed brackets : An in vitro study.
J Orofac Orthop 2024;
85:327-339. [PMID:
36862188 PMCID:
PMC11358261 DOI:
10.1007/s00056-023-00460-8]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 05/24/2022] [Accepted: 12/29/2022] [Indexed: 03/03/2023]
Abstract
PURPOSE
To investigate a novel in-office three-dimensionally (3D) printed polymer bracket regarding slot precision and torque transmission.
METHODS
Based on a 0.022″ bracket system, stereolithography was used to manufacture brackets (N = 30) from a high-performance polymer that met Medical Device Regulation (MDR) IIa requirements. Conventional metal and ceramic brackets were used for comparison. Slot precision was determined using calibrated plug gages. Torque transmission was measured after artificial aging. Palatal and vestibular crown torques were measured from 0 to 20° using titanium-molybdenum (T) and stainless steel (S) wires (0.019″ × 0.025″) in a biomechanical experimental setup. The Kruskal-Wallis test with post hoc test (Dunn-Bonferroni) was used for statistical analyses (significance level p < 0.05).
RESULTS
The slot sizes of all three bracket groups were within the tolerance range according to DIN 13996 (ceramic [C]: 0.581 ± 0.003 mm; metal [M]: 0.6 ± 0.005 mm; polymer [P]: 0.581 ± 0.010 mm). The maximum torque values of all bracket-arch combinations were above the clinically relevant range of 5-20 Nmm (PS: 30 ± 8.6 Nmm; PT: 27.8 ± 14.2 Nmm; CS: 24 ± 5.6 Nmm; CT: 19.9 ± 3.8 Nmm; MS: 21.4 ± 6.7 Nmm; MT: 16.7 ± 4.6 Nmm).
CONCLUSIONS
The novel, in-office manufactured polymer bracket showed comparable results to established bracket materials regarding slot precision and torque transmission. Given its high individualization possibilities as well as enabling an entire in-house supply chain, the novel polymer brackets bear high potential of future usage for orthodontic appliances.
Collapse