Menchini M, Sartini F, Tatti F, Peiretti E, Figus M. Endothelial cell density change in fully dislocated XEN Gel IMPLANT after trabeculectomy: A case report.
Eur J Ophthalmol 2021;
32:NP14-NP18. [PMID:
33472434 DOI:
10.1177/1120672121989632]
[Citation(s) in RCA: 2] [Impact Index Per Article: 0.7] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/16/2022]
Abstract
PURPOSE
To report a case of fully dislocated XEN Gel Implant device into the anterior chamber, 18 months after its implantation and its impact on endothelial cell density.
RESULT
A 75-year-old man with a history of multiple glaucoma surgeries during the last three years presented with an entirely dislocated XEN Gel implant into the anterior chamber. An endothelial cell density reduction was observed over 18 months. Thus, the implant was removed. In the short-term after surgery, the endothelial cell loss rate reduced.
CONCLUSION
XEN Gel Implant can dislocate into the anterior chamber, increasing endothelial cell loss in an eye already underwent multiple surgical procedures with subsequent adverse events. Therefore, if the implant seems displaced, the endothelial cell density should be monitored, and the length of the free-tube segment within the anterior chamber should be measured, during the patient's follow-up.