Swinnen J, Fletcher JP, Wong KP, Young N, Simmons K. EVT endovascular graft for abdominal aortic aneurysm.
ANZ J Surg 2001;
71:403-6. [PMID:
11450914 DOI:
10.1046/j.1440-1622.2001.02145.x]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/20/2022]
Abstract
BACKGROUND
A variety of prostheses are now available for the endovascular treatment of abdominal aortic aneurysm (AAA). Significant advantages of the EVT device are its unibody design, secure hook attachment system and graft fabric approximating that used in conventional surgery.
METHODS
Implantation of the EVT device was attempted in 60 patients who were studied prospectively with an analysis of subsequent problems encountered.
RESULTS
Conversion to open repair was required in four cases (6.7%). There were nine tube grafts inserted, 13 aorto-unilateral iliac with crossover grafts and 34 aorto-bi-iliac grafts. There was one death (mortality 1.7%). Endoleaks were identified in eight patients (14%), none of which were proximal; three sealed spontaneously, two were treated with coil embolization, two are being observed and one patient had an iliac attachment converted to an open anastomosis. Access vessel problems were seen in 21 patients (35%); two-thirds were corrected at the time of initial surgery. Seven patients (12%) had primary graft limb problems identified and treated before leaving the operating room. Nine patients (16%) developed secondary graft limb problems, which were diagnosed and treated after the initial surgery. Endovascular treatment was used in eight and was successful in six with surgical revision required in two. On review of these cases to assess if the problem could have been predicted at the time of initial surgery, it was felt that more aggressive treatment of intraoperatively diagnosed graft limb stenoses, even though considered mild, may have prevented 50% of subsequent secondary graft limb occlusions.
CONCLUSION
Although the EVT device has significant advantages in the endovascular management of aortic aneurysm, potential graft limb problems need to be actively identified with the majority able to be successfully managed by supplementary endovascular techniques.
Collapse