van Schilt KLJ, Hollander EJF, Koelemay MJ, van Geloven AAW, Olthof DC. Assessment of microcirculatory changes in local tissue oxygenation after revascularization for peripheral arterial disease with the Hyperview®, a portable hyperspectral imaging device.
Vascular 2023;
31:961-967. [PMID:
35580351 DOI:
10.1177/17085381221102813]
[Citation(s) in RCA: 1] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/17/2022]
Abstract
OBJECTIVES
The Hyperview® is a hyperspectral camera, which can be used to assess the microcirculation of patients with peripheral arterial disease (PAD) and/or diabetes mellitus (DM). It measures local tissue oxygenation in concentrations of oxyhemoblobin (OXY), deoxyhemoglobin (DEOXY), and O2-saturation (O2-SAT) in arbitrary units. The aim of this validation study is to assess whether the Hyperview® is able to monitor microcirculatory changes after revascularization in patients with PAD.
METHODS
In this prospective observational cohort study, 50 patients with PAD were included who were scheduled for endovascular, hybrid, or open revascularization. The ankle-brachial index (ABI), systolic toe pressure (TP) (in case of DM), and a set of Hyperview® measurements of the plantar region were recorded before and after treatment. Changes in pre- and postoperative measurements were assessed with the paired t-test.
RESULTS
Some 38 patients underwent endovascular revascularization and eight patients underwent hybrid or open vascular surgical revascularization. After revascularization, the ABI improved from 0.58 to 0.80 (p < 0.001). OXY increased from 72.6 to 77.8 (p = 0.134). DEOXY decreased from 69.1 to 55.0 (p < 0.001). O2-SAT increased from 51.3 to 58.1 (p < 0.001). TP improved from 50.6 to 61.9 mmHg (p = 0.065) but was measured in only 16 patients.
CONCLUSION
The Hyperview® is able to observe changes in the microcirculation after revascularization in terms of DEOXY and O2-SAT. The results of this study are a promising step into the validation of the Hyperview®.
Collapse