Inuzuka R, Tachimori H, Kim SH, Matsui H, Kobayashi T, Kato A, Fujii T, Ho M, Morikawa H, Takahashi S, Shirato H, Haishima Y, Okamoto Y, Sakoda H, Tomita H. Practice and Safety of Static Balloon Atrial Septostomy Based on a Nationwide Registry Data.
Circ J 2022;
86:1990-1997. [PMID:
36047087 DOI:
10.1253/circj.cj-22-0185]
[Citation(s) in RCA: 2] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Indexed: 09/06/2024]
Abstract
BACKGROUND
Balloon atrial septostomy (BAS) is an essential catheterization procedure for congenital heart lesions. Recently, a balloon catheter for static BAS was approved for the first time in Japan as an alternative to the conventional pull-through BAS. Despite the expected increase in the use of static BAS, reports on its safety are scarce worldwide.
METHODS AND RESULTS
Data on static and pull-through BAS registered in a national registry between 2016 and 2018 were collected. During the study period, 247 sessions of static BAS and 588 sessions of pull-through BAS were performed on a total of 674 patients. Patients who underwent static BAS were older (P<0.001). The incidence of serious adverse events (4.3% vs. 0.9%, P=0.03) and the overall incidence of adverse events (8.1% vs. 3.2%, P=0.03) were higher in static BAS than in pull-through BAS. Among patients who underwent static BAS, the risk factor for adverse events was a body weight <3 kg at the time of the procedure (odds ratio: 4.3 [confidence interval: 1.7-11], P=0.003).
CONCLUSIONS
This nationwide study revealed differences in patient background between static and pull-through BAS, as well as a higher incidence of adverse events related to static BAS. Patients weighing <3 kg are at high risk for adverse events after static BAS and may require surgical and circulatory support backup.