De Cooman S, Hendrickx JFA, Peyton PJ, Demeere JL, De Wolf AM. Agent consumption with the Zeus® in the automated closed circuit anesthesia mode with O2/air mixtures.
BMC Res Notes 2014;
7:469. [PMID:
25056252 PMCID:
PMC4118792 DOI:
10.1186/1756-0500-7-469]
[Citation(s) in RCA: 4] [Impact Index Per Article: 0.4] [Reference Citation Analysis] [Abstract] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 01/04/2014] [Accepted: 07/15/2014] [Indexed: 11/10/2022] Open
Abstract
BACKGROUND
Earlier software versions of the Zeus® (Lübeck, Dräger, Germany) failed to provide true closed circuit anesthesia (CCA) conditions. We examined whether the latest software (SW 4.03 MK 04672-00) achieves this goal.
METHODS
In 8 ASA I-III patients, the CCA mode of the Zeus® was used to maintain the inspired O2 (FIO2) and end-expired sevoflurane % (FAsevo) at 50 and 1.8%, respectively. The fresh gas flow (FGF) of O2 and air and the sevoflurane injection rate (=Vinjsevo, mL liquid sevo/h) were videotaped from the control screen and entered offline into a spreadsheet. Cumulative sevoflurane usage during early wash-in (=0-1 min, CDsevo0-1), late wash-in (=1-5 min, CDsevo1-5), and maintenance (=5-60 min, CDsevo5-60) was calculated, and Vinjsevo between 1 and 60 min was compared with published uptake data.
RESULTS
FAsevo reached 1.8% within 101 (23) sec. CDsevo0-1 was between 1.24 (0.03) and 3.01(0.25) mL (a range is provided because no absolute Vinjsevo values were displayed once Vinjsevo was > 100 mL/h, which occurred between 15 ± 2 and 46 ± 6 sec). CDsevo1-5 was 0.81 (0.37) mL, and CDsevo5-60 was 4.63 (0.94) mL. The Vinjsevo pattern between 1 and 60 min matched previously published uptake data. Brief high FGF periods were used to maintain the target FIO2, and to refill the reservoir bag after external pressure had been applied to the abdomen; subsequent "spikes" wasted 0.08-0.19 mL and 0.14-0.49 mL sevoflurane (1-3% and 3-9% of total agent usage between 1 and 60 min, respectively).
CONCLUSION
Under the conditions specified, the Zeus® approaches CCA conditions so closely that further reductions in agent usage would have minimal economic significance.