Haenen FW, Van Oostende C, Allegaert M, Round KJ, Rosen JL, Guy ST, Rodrigus I. Prosthesis-prosthesis anastomosis using barbed sutures compared to conventional sutures under high, long-term pressure; in vitro extracorporeal circulation setup.
Perfusion 2024;
39:571-577. [PMID:
36691745 DOI:
10.1177/02676591231153539]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 01/25/2023]
Abstract
BACKGROUND
While barbed sutures have been extensively utilized in other disciplines, they have not been widely adopted in cardiac surgery. The lack of safety and feasibility data has limited its use within the field. To aide in the further understanding of how cardiac surgeons can use barbed sutures, we sought to develop a high-pressure in vitro simulation model. We compared knotless barbed sutures in a highly pressurized anastomosis to conventional sutures.
METHODS
Ten specimens in total were utilized in prosthesis anastomosis, using 34 mm Gelweave Plexus (Terumo Aortic, Sunrise, FL 33325, USA) and 34 mm Hemabridge (Intergard Woven Hemabridge, Getinge, Göteborg, Sweden). Five models of size 3-0 barbed suture anastomoses using non-absorbable, barbed, self-retaining, monofilament polypropylene sutures (Filbloc® 3-0, Assut Europe, Rome, Italy) were compared against five conventional anastomoses using size 4-0 polypropylene monofilament (Ethicon, USA). The systems were connected using a novel-designed extracorporeal circulation system. Pressure was rapidly increased in the specimen to a mean pressure of 300-350 mmHg, running then for a minimum of 48 hours to assess anastomosis strength and endurance.
RESULTS
No anastomotic dehiscence or rupture was recorded. Complex, angular anastomosis required extra stitch leakage sutures in both conventional and barbed suture specimens.
CONCLUSION
Using knotless barbed sutures with an additional self-locking maneuver for prosthesis-prosthesis anastomosis in cardiac surgery is feasible in an in vitro model under long term, high-mean pressure when compared to conventional sutures. In vivo trials should be performed to further validate the in vitro findings.