Massmann A, Fries P, Shayesteh-Kheslat R, Buecker A, Berg P, Frenzel F. Feasibility of aortic aneurysm sac embolization using a novel shape memory polymer embolic device.
Eur Radiol Exp 2023;
7:12. [PMID:
37009937 PMCID:
PMC10068687 DOI:
10.1186/s41747-023-00328-x]
[Citation(s) in RCA: 2] [Impact Index Per Article: 2.0] [Reference Citation Analysis] [Abstract] [Key Words] [Journal Information] [Subscribe] [Scholar Register] [Received: 07/19/2022] [Accepted: 01/30/2023] [Indexed: 04/04/2023] Open
Abstract
BACKGROUND
We investigated the feasibility of aneurysm sac embolization using a novel self-expanding porous shape memory polymer (SMP) device during endovascular aortic abdominal or thoracic aneurysm repair (EVAR).
METHODS
Retrospective analysis of consecutive patients treated at 2 centers in Germany. Patients were treated from January 2019 to July 2021 with follow-up at 7 days and 3, 6, and 12 months. Aneurysm sacs were implanted with SMP devices immediately following endograft placement during the same procedure. Primary endpoint was technically successful SMP-device deployment into the aneurysm sac outside the endograft. Secondary endpoints were changes in aneurysm volume and associated complications (e.g., endoleaks).
RESULTS
We included 18 patients (16 males), aged 72 ± 9 years, achieving 100% technical success. Mean preprocedure aortic aneurysm sac volume was 195 ± 117 mL with a perfused aneurysm volume of 97 ± 60 mL. A mean of 24 ± 12 SMP devices per patient were used (range 5-45, corresponding to 6.25-56.25 mL expanded embolic material volume). All evaluable patients exhibited sac regression except 2 patients yet to reach 3-month follow-up. At mean 11 ± 7 months (range 3-24), change in aneurysm volume from baseline was -30 ± 21 mL (p < 0.001). In 8 patients, aneurysm regression was observed despite type 2 endoleaks in 6 and type 1A endoleaks in 2, none of them requiring further intervention to date. No morbidity or mortality related to this treatment occurred.
CONCLUSIONS
SMP devices for aortic aneurysm sac embolization during endovascular repair appear feasible and safe in this small case series. Prospective studies are needed.
KEY POINTS
• Shape memory polymer is a novel, self-expanding, porous, and radiolucent embolic device material. • Aortic aneurysm sacs were treated with polymer devices immediately following endograft placement. • Aortic aneurysm sac regression was observed in all patients with over 3-month follow-up. • Aortic aneurysm sac regression was observed even in the presence of endoleaks.
Collapse