Yuasa S, Masuyama S, Soeda T, Matsuda M, Shirota K, Taira H. Surgical removal of an accufix pacing lead with a protruding J wire.
THE JAPANESE JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY : OFFICIAL PUBLICATION OF THE JAPANESE ASSOCIATION FOR THORACIC SURGERY = NIHON KYOBU GEKA GAKKAI ZASSHI 2002;
50:443-4. [PMID:
12428386 DOI:
10.1007/bf02913180]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 11/25/2022]
Abstract
We report a patient with a fractured J wire protruding through the outer polyurethane sheath of an Accufix electrode in the subclavian vein and right atrium. The wire within the subclavian vein was removed transvenously, while the tip of the lead within the right atrium was removed surgically via a median sternotomy.