Automatic validation of the gantry tilt in a computed tomography scanner using a head polymethyl methacrylate phantom. POLISH JOURNAL OF MEDICAL PHYSICS AND ENGINEERING 2021. [DOI: 10.2478/pjmpe-2021-0008] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/20/2022]
Abstract
Abstract
The purpose of this study was to develop an automatic method for validating the computed tomography gantry tilt. A head polymethyl methacrylate phantom with a diameter of 16 cm was used. Gantry tilt angles were measured both manually and automatically. Manual measurements were performed by measuring the length of the anteroposterior and lateral diameters from acquired images using electronic calipers. Automatic measurements consisted of a number of steps: phantom segmentation, determination of the center of the phantom, measurement of the anteroposterior and lateral diameters, and computation of the gantry tilt angle. The method was implemented on the gantry angles from 0° to 15°. The proposed method of measuring gantry angles produced accurate gantry tilt angles. The differences with the angles displayed on the gantry were less than 1°. The results of the automatic method were the same as those of the manual method (R2 > 0.98).
Collapse
|
2
|
|
3
|
Cave EM, Weston MJ. Comparison of light reflection rheography with ascending venography in the diagnosis of lower limb venous thrombosis. Br J Radiol 1995; 68:1136-7. [PMID: 7496724 DOI: 10.1259/0007-1285-68-814-1136-b] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 01/25/2023] Open
|