Aranda E, Iha S, Solari S, Rodríguez D, Romero V, Villarroel L, Pereira J, Panes O, Mezzano D. Serotonin secretion by blood platelets: accuracy of high-performance liquid chromatography-electrochemical technique compared with the isotopic test and use in a clinical laboratory.
Res Pract Thromb Haemost 2023;
7:102156. [PMID:
37601022 PMCID:
PMC10439442 DOI:
10.1016/j.rpth.2023.102156]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 02/14/2023] [Revised: 06/07/2023] [Accepted: 07/06/2023] [Indexed: 08/22/2023] Open
Abstract
Background
Mild secretion defects are the most frequent and challenging blood platelet disorders to diagnose. Most δ-granule secretion tests lack validation, are not quantitative, or have unreliable response to weak platelet agonists.
Objectives
To compare platelet serotonin secretion by HPLC-electrochemical detection technique (HPLC-ECD) with the reference isotopic test (3H-5-HT), evaluating its performance in clinical laboratories.
Methods
The assay validation followed STARD-2015 recommendations. HPLC-ECD measured the nonsecreted serotonin remaining in platelet pellets after aggregation, comparing it with the reference 3H-5-HT assay. We studied subjects with inherited and aspirin-induced blood platelet disorders and assessed the HPLC-ECD operation for routine clinical diagnosis.
Results
Calibration curves were linear (R2 = 0.997), with SD for residuals of 3.91% and analytical sensitivity of 5ng/mL. Intra- and interassay imprecision bias ranged between -8.5% and 2.1% and -9% and 3.1%, respectively. Serotonin recovery and stability were >95%, and the variability range of measurements was -5.5% to 4.6%. Statistical differences detected between tests were biologically irrelevant, with bias of 1.48% (SD, 8.43) and CI agreement of -18% to 15%. Both assays distinctly detected platelet secretion induced by 10 μM epinephrine and 4 μmM adenosine diphosphate. However, HPLC-ECD is quantitative and more sensitive to low serotonin content in blood platelets. Reference cutoffs for each agonist were determined in 87 subjects. Initially, the HPLC-ECD requires relatively expensive equipment and trained operators but has remarkably cheap running costs and a turn-around time of 24-36 hours. We have used this diagnostic tool routinely for >8 years.
Conclusion
HPLC-ECD assay for platelet serotonin secretion is highly accurate, has advantages over the reference 3H-5-HT test, and is suitable as a clinical laboratory technique.