Yang Y, Ning R, Xu S, Xiahou J, Li J, Savini G, Schiano-Lomoriello D, Zhou X, Huang J. Evaluation of the Agreement Between a New Pyramid Wavefront Sensor Aberrometer and Scheiner-Smirnov Aberrometers.
J Refract Surg 2024;
40:e218-e228. [PMID:
38593257 DOI:
10.3928/1081597x-20240311-02]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 04/11/2024]
Abstract
PURPOSE
To assess agreement between a new aberrometer (Osiris-T; CSO) employing pyramid wavefront sensor technique and Scheiner-Smirnov aberrometer (OPD-Scan III; Nidek) on measuring ocular, corneal, and internal aberrations in healthy participants.
METHODS
The measurements were conducted three times consecutively by an experienced examiner. The total root mean square (RMS) aberrations, higher order aberration RMS, coma Z3±1, trefoil Z3±3, spherical aberration Z40, and astigmatism II Z4±2 up to 7th order were exported in both 4-and 6-mm pupil zones. The parameters between the two devices were statistically compared using the paired t-test, and the differences assessed with Bland-Altman plots and 95% limits of agreement.
RESULTS
This prospective study included 70 right eyes of 70 healthy participants with an average age of 25.94 ± 6.59 years (range: 18 to 47 years). The mean difference in the two devices ranged from 0.01 µm for astigmatism II Z4±2 to 0.63 µm for total RMS in 4 mm and from 0.01 to 1.41 µm in 6-mm pupil size. The Bland-Altman analysis of ocular, corneal, and internal aberrations indicated high agreement between the two devices and the maximum absolute values for 95% limits of agreement ranged from 0.03 to 1.06 µm for 4-mm pupil diameters and 0.12 to 1.13 µm for 6-mm pupil diameters.
CONCLUSIONS
The newly developed pyramid wavefront sensor technique aberrometer demonstrated a high agreement with a Scheiner-Smirnov aberrometer when measuring ocular, corneal, and internal aberrations in healthy participants. Thus, the two aberrometers may be considered interchangeable for clinical applications. [J Refract Surg. 2024;40(4):e218-e228.].