Sun S, Sun M, Tang J, Yang F, Liu Z, Zhao S, Huang Y. A comparative study of pupil offset measurement using Pentacam and Keratron Scout in myopic young adults.
Clin Exp Optom 2024;
107:40-46. [PMID:
37156100 DOI:
10.1080/08164622.2023.2203316]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 02/09/2023] [Accepted: 04/11/2023] [Indexed: 05/10/2023] Open
Abstract
CLINICAL RELEVANCE
Assessing the consistency of pupil offset measurements between the Pentacam and Keratron Scout is essential for the refractive surgery design of young myopic patients.
BACKGROUND
Accurate preoperative measurement of pupil offset is critical to achieving a better quality of vision after refractive surgery. The Pentacam and Keratron Scout are commonly used in hospitals, and evaluating their consistency is essential for accurate measurement of pupil offset.
METHODS
Six hundred eyes (600 subjects) were included in this study. Pupil offset and its X, Y-component were measured by Pentacam and Keratron Scout, respectively. Agreement and repeatability between the two devices were identified by intraclass correlation coefficient and Bland-Altman plots with 95% limits of agreement. Paired t-tests and Pearson analysis were used to compare the differences and correlations between the two devices.
RESULTS
The mean age of all subjects was 23 ± 5 years. The mean pupil offset magnitude obtained from Pentacam and Keratron Scout was 0.16 ± 0.08 mm and 0.15 ± 0.07 mm. The 95% limits of agreement (-0.11-0.13, -0.09-0.11, and -0.11-0.12) and intraclass correlation coefficient (0.82, 0.84, and 0.81) demonstrated good agreement and repeatability of the two devices in measuring pupil offset and its X, Y-component. A significant correlation between the two devices was found (r = 0.71, 0.73, and 0.70). The direction of pupil offset measured by the devices was both predominately towards the superonasal quadrant.
CONCLUSION
Pentacam and Keratron Scout showed good agreement in measuring pupil offset and its X, Y-component, which can be used interchangeably in clinical practice.