Fan X, Li Q, Fang H, Ju Y, Jin Z, Li H, Zhang X. Development and application of a high accuracy method for measuring Pb in blood.
Clin Chim Acta 2023;
538:164-168. [PMID:
36423701 DOI:
10.1016/j.cca.2022.11.014]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 09/05/2022] [Revised: 10/17/2022] [Accepted: 11/08/2022] [Indexed: 11/23/2022]
Abstract
BACKGROUND
Lead (Pb) is an environmental pollutant, and human exposure is assessed by measuring blood Pb concentrations. Today, the use of Pb-based products is restricted, but the health effects of low Pb concentrations require investigation. For this, a high-accuracy method is needed. Here, we report a new inductively coupled plasma mass spectrometry (ICP-MS) method for quantifying Pb in blood.
METHODS
Blood samples and calibrators were gravimetrically spiked with bismuth as an internal standard, digested with ultrapure nitric acid, and diluted 100 times. The calibrators were made from high-purity reference materials supplied by the National Institute of Standards and Technology, and the sample concentration was measured using the calibration bracketing method. The analytical performance and application of the new method were investigated.
RESULTS
The method had good specificity and a negligible matrix effect. The intra- and inter-assay coefficients of variation (CVs) were <1.34 % and <1.88 %, respectively, and the total CV was <1.74 %. The range of recovery was 98.9 %-99.4 %. SRM955d, BCR-635, and BCR-636 were within the certified intervals, with mean relative bias values of 0.93 %, -0.13 %, and -0.56 %, respectively. The limit of quantification was 1.1 µg/l. The method was used to assign target values to external quality assessment samples and to assess routine methods, showing good results at high concentrations. However, at low concentrations, two laboratories did not pass.
CONCLUSIONS
This ICP-MS method is a simple, accurate method and can be a candidate reference method for blood Pb measurement. But more research will be needed to verify this.