Asif M, Groboske SE, Leung EKY, Yeo KTJ, van Wijk XMR. Evaluation of a New Generation Automated Assay for 25-Hydroxy Vitamin D Based on Competitive Protein Binding.
J Appl Lab Med 2019;
4:247-253. [PMID:
31639671 DOI:
10.1373/jalm.2018.028415]
[Citation(s) in RCA: 5] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 10/13/2018] [Accepted: 01/22/2019] [Indexed: 11/06/2022]
Abstract
BACKGROUND
The interest for vitamin D has exponentially increased testing demand for 25-hydroxy vitamin D [25(OH)D]. Consequently, many laboratories are switching from LC-MS/MS methods to automated, high-throughput immunoassays. One of the major potential issues with these assays has been the lack of cross-reactivity with 25(OH)D2.
METHODS
We have evaluated the Roche Elecsys vitamin D total II assay for accuracy by comparing 79 patient samples with LC-MS/MS. The cross-reactivity for 25(OH)D2 was evaluated by analyzing samples with high 25(OH)D2 separately and estimating 25(OH)D2 recovery, as well as by spiking of 25(OH)D2. The assay was further evaluated for precision, linearity, sample type, and common interferences.
RESULTS
There was mostly good agreement between the Elecsys and LC-MS/MS assays (Deming regression: y = 0.95x + 0.70), with an overall bias of 2.3% (-0.84 ng/mL). However, there were 6 out of 79 (7.6%) discordant samples. The Deming regression for samples with high 25(OH)D2 compared to LC-MS/MS showed similar slope and intercept (y = 0.97x - 1.1). The average recovery of 25(OH)D2 for these samples was 90%. The initial precision studies were in general agreement with the package insert, but long-term clinical use showed higher-than-claimed imprecision (11.7%-14.4% at 12 ng/mL and 6.9%-7.6% at 27 ng/mL; claimed: 7.2% and 5.0%, respectively). We observed 1 falsely high result in plasma, an issue previously addressed by Roche in a medical device correction.
CONCLUSIONS
The analytical performance of the Roche Vitamin D assay was acceptable, and the assay had a good cross-reactivity for 25(OH)D2.