Hurwitz BJ, Zion MM, Obel IW. Myocardial perforation by flexible Elema endocardial pacing catheters.
Thorax 1974;
29:678-84. [PMID:
4450177 PMCID:
PMC470223 DOI:
10.1136/thx.29.6.678]
[Citation(s) in RCA: 2] [Impact Index Per Article: 0.0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 01/10/2023]
Abstract
Hurwitz, B. J., Zion, M. M., and Promund Obel, I. W. (1974).Thorax, 29, 678-684. Myocardial perforation by flexible Elema endocardial pacing catheters. Myocardial perforation as a complication of permanent endocardial pacing with flexible Elema catheters is described in nine patients. In eight patients pacing failed, but there were no other serious ill-effects. Diaphragmatic pacing occurred in five patients. Radiological evidence of movement, either posteriorly and/or laterally, of the electrode tip was detected in five patients. In only two of them was the electrode seen to be definitely extracardiac in position. A change in the pacing wave form aided the diagnosis in two patients. In a further two patients electrode tip electrocardiograms helped to confirm the diagnosis where, upon withdrawal of the perforating electrode, small Q waves disappeared, RS complexes enlarged, and the ST segments became more elevated, with deeply inverted T waves.
Withdrawal and repositioning of the electrode catheters was performed in seven patients. In one, a new catheter was inserted, and in another, where uninterrupted cardiac pacing was accompanied by intermittent diaphragmatic pacing, the pacing system was left unaltered. The follow-up is from six months to four and a half years with one late death apparently unrelated to pacing failure.
Collapse