1
|
van den Besselaar AMHP, Stavelin A, Kitchen S, Bryant M, Tripodi A, Scalambrino E, Clerici M, Herbel P, Jünschke A, Meyer Dos Santos S, Meijer P, Niessen RWLM, Meijers JCM, Thelwell C, Cuker A, Kung C, Cao Z, Zander N, Iwasaki Y, Depasse F, van Rijn C, Baktawar S, Abdoel C, Cobbaert CM. Defining a metrologically traceable and sustainable calibration hierarchy of international normalized ratio for monitoring of vitamin K antagonist treatment in accordance with International Organization for Standardization (ISO) 17511:2020 standard: communication from the International Federation of Clinical Chemistry and Laboratory Medicine-SSC/ISTH working group on prothrombin time/international normalized ratio standardization. J Thromb Haemost 2024; 22:1236-1248. [PMID: 38128762 DOI: 10.1016/j.jtha.2023.12.013] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 09/18/2023] [Revised: 12/01/2023] [Accepted: 12/07/2023] [Indexed: 12/23/2023]
Abstract
Calibration of prothrombin time (PT) in terms of international normalized ratio (INR) has been outlined in "Guidelines for thromboplastins and plasmas used to control oral anticoagulant therapy" (World Health Organization, 2013). The international standard ISO 17511:2020 presents requirements for manufacturers of in vitro diagnostic (IVD) medical devices (MDs) for documenting the calibration hierarchy for a measured quantity in human samples using a specified IVD MD. The objective of this article is to define an unequivocal, metrologically traceable calibration hierarchy for the INR measured in plasma as well as in whole blood samples. Calibration of PT and INR for IVD MDs according to World Health Organization guidelines is similar to that in cases where there is a reference measurement procedure that defines the measurand for value assignment as described in ISO 17511:2020. We conclude that, for PT/INR standardization, the optimal calibration hierarchy includes a primary process to prepare an international reference reagent and measurement procedure that defines the measurand by a value assignment protocol conforming to clause 5.3 of ISO 17511:2020. A panel of freshly prepared human plasma samples from healthy adult individuals and patients on vitamin K antagonists is used as a commutable secondary calibrator as described in ISO 17511:2020. A sustainable metrologically traceable calibration hierarchy for INR should be based on an international protocol for value assignment with a single primary reference thromboplastin and the harmonized manual tilt tube technique for clotting time determination. The primary international reference thromboplastin reagent should be used only for calibration of successive batches of the secondary reference thromboplastin reagent.
Collapse
Affiliation(s)
- Antonius M H P van den Besselaar
- Coagulation Reference Laboratory, Department of Clinical Chemistry and Laboratory Medicine, Leiden University Medical Center, Leiden, the Netherlands.
| | - Anne Stavelin
- The Norwegian Organisation for Quality Improvement of Laboratory Examinations, Bergen, Norway
| | - Steve Kitchen
- Sheffield Haemophilia and Thrombosis Centre, Royal Hallamshire Hospital, Sheffield, UK
| | - Michelle Bryant
- Sheffield Haemophilia and Thrombosis Centre, Royal Hallamshire Hospital, Sheffield, UK
| | - Armando Tripodi
- Fondazione Istituto di Ricovero e Cura a Carattere Scientifico Ca' Granda Ospedale Maggiore Policlinico, Angelo Bianchi Bonomi Hemophilia and Thrombosis Center, Milano and Fondazione Luigi Villa, Milano, Italy
| | - Erica Scalambrino
- Fondazione Istituto di Ricovero e Cura a Carattere Scientifico Ca' Granda Ospedale Maggiore Policlinico, Angelo Bianchi Bonomi Hemophilia and Thrombosis Center, Milano and Fondazione Luigi Villa, Milano, Italy
| | - Marigrazia Clerici
- Fondazione Istituto di Ricovero e Cura a Carattere Scientifico Ca' Granda Ospedale Maggiore Policlinico, Angelo Bianchi Bonomi Hemophilia and Thrombosis Center, Milano and Fondazione Luigi Villa, Milano, Italy
| | | | | | | | - Piet Meijer
- External quality Control of diagnostic Assays and Tests (ECAT) Foundation, Voorschoten, the Netherlands
| | | | - Joost C M Meijers
- Department of Molecular Hematology, Sanquin Research, Amsterdam, the Netherlands; Department of Experimental Vascular Medicine, Amsterdam Cardiovascular Sciences, Amsterdam University Medical Center, University of Amsterdam, Amsterdam, the Netherlands
| | - Craig Thelwell
- Therapeutic Reference Materials, Medicines and Healthcare Products Regulatory Agency, National Institute for Biological Standards and Control, Potters Bar, UK
| | - Adam Cuker
- Department of Medicine, Perelman School of Medicine, University of Pennsylvania, Philadelphia, Pennsylvania, USA; Department of Pathology and Laboratory Medicine, Perelman School of Medicine, University of Pennsylvania, Philadelphia, Pennsylvania, USA
| | | | | | | | | | | | - Claudia van Rijn
- Coagulation Reference Laboratory, Department of Clinical Chemistry and Laboratory Medicine, Leiden University Medical Center, Leiden, the Netherlands
| | - Shanti Baktawar
- Coagulation Reference Laboratory, Department of Clinical Chemistry and Laboratory Medicine, Leiden University Medical Center, Leiden, the Netherlands
| | - Charmane Abdoel
- Coagulation Reference Laboratory, Department of Clinical Chemistry and Laboratory Medicine, Leiden University Medical Center, Leiden, the Netherlands
| | - Christa M Cobbaert
- Coagulation Reference Laboratory, Department of Clinical Chemistry and Laboratory Medicine, Leiden University Medical Center, Leiden, the Netherlands
| |
Collapse
|
2
|
Loh TP, Markus C, Tan CH, Tran MTC, Sethi SK, Lim CY. Lot-to-lot variation and verification. Clin Chem Lab Med 2022; 61:769-776. [PMID: 36420533 DOI: 10.1515/cclm-2022-1126] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 11/07/2022] [Accepted: 11/14/2022] [Indexed: 11/24/2022]
Abstract
Abstract
Lot-to-lot verification is an integral component for monitoring the long-term stability of a measurement procedure. The practice is challenged by the resource requirements as well as uncertainty surrounding experimental design and statistical analysis that is optimal for individual laboratories, although guidance is becoming increasingly available. Collaborative verification efforts as well as application of patient-based monitoring are likely to further improve identification of any differences in performance in a relatively timely manner. Appropriate follow up actions of failed lot-to-lot verification is required and must balance potential disruptions to clinical services provided by the laboratory. Manufacturers need to increase transparency surrounding release criteria and work closer with laboratory professionals to ensure acceptable reagent lots are released to end users. A tripartite collaboration between regulatory bodies, manufacturers, and laboratory medicine professional bodies is key to developing a balanced system where regulatory, manufacturing, and clinical requirements of laboratory testing are met, to minimize differences between reagent lots and ensure patient safety. Clinical Chemistry and Laboratory Medicine has served as a fertile platform for advancing the discussion and practice of lot-to-lot verification in the past 60 years and will continue to be an advocate of this important topic for many more years to come.
Collapse
Affiliation(s)
- Tze Ping Loh
- Department of Laboratory Medicine , National University Hospital , Singapore , Singapore
| | - Corey Markus
- Flinders University International Centre for Point-of-Care Testing , Flinders Health and Medical Research Institute , Adelaide , SA , Australia
| | - Chin Hon Tan
- Department of Industrial and Systems Engineering , National University of Singapore , Singapore , Singapore
| | - Mai Thi Chi Tran
- Faculty of Medical Technology , Hanoi Medical University , Hanoi , Vietnam
- Department of Clinical Biochemistry , National Children’s Hospital , Hanoi , Vietnam
| | - Sunil Kumar Sethi
- Department of Laboratory Medicine , National University Hospital , Singapore , Singapore
| | - Chun Yee Lim
- Engineering Cluster , Singapore Institute of Technology , Singapore , Singapore
| |
Collapse
|
3
|
van den Besselaar AMHP, van Rijn CJJ, Abdoel CF, Cobbaert CM. Bias and uncertainty of the International Normalized Ratio determined with a whole blood point-of-care prothrombin time test device by comparison to a new International Standard for thromboplastin. Thromb Res 2021; 202:1-7. [PMID: 33706049 DOI: 10.1016/j.thromres.2021.02.018] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 10/26/2020] [Revised: 02/04/2021] [Accepted: 02/10/2021] [Indexed: 10/22/2022]
Abstract
BACKGROUND Whole blood point-of-care PT/INR test devices, e.g. CoaguChek XS, are calibrated by their manufacturers. In the Netherlands, each new lot of test strips for CoaguChek XS is validated by a group of anticoagulant clinics collaborating with a Coagulation Reference Laboratory. In 2017, a new International Standard for recombinant human thromboplastin (coded rTF/16) has been established by the World Health Organization. AIM To assess uncertainty of the validation procedure and the magnitude of the INR bias of a series of consecutive lots of test strips imported in the Netherlands. METHODS CoaguChek XS test strip INR results were compared to INRs determined with the new International Standard rTF/16. Comparisons were made with variable numbers of blood samples obtained from patients treated with vitamin K-antagonists. Relationships between CoaguChek XS and rTF/16 results were determined with orthogonal regression analysis. The relationships were used to assess bias and uncertainty of bias. RESULTS Average bias between CoaguChek XS test results and rTF/16 depends on the INR level. Overall, there was a trend of increasing bias and increasing uncertainty with increasing INR values. Along the sequence of 47 consecutive lots, a temporary fluctuation of bias was observed. At an INR level of 3.0 the average bias was less than 10% in all cases, but at an INR of 4.0 there were 5 lots with average bias between 10 and 15%. CONCLUSION Validation of test strips is useful to assess bias but depends on availability of fresh patients' samples and traceability to an accepted Reference Measurement System.
Collapse
Affiliation(s)
- Antonius M H P van den Besselaar
- Coagulation Reference Laboratory, Department of Clinical Chemistry and Laboratory Medicine, Leiden University Medical Center, Leiden, the Netherlands.
| | - Claudia J J van Rijn
- Coagulation Reference Laboratory, Department of Clinical Chemistry and Laboratory Medicine, Leiden University Medical Center, Leiden, the Netherlands
| | - Charmane F Abdoel
- Coagulation Reference Laboratory, Department of Clinical Chemistry and Laboratory Medicine, Leiden University Medical Center, Leiden, the Netherlands
| | - Christa M Cobbaert
- Coagulation Reference Laboratory, Department of Clinical Chemistry and Laboratory Medicine, Leiden University Medical Center, Leiden, the Netherlands
| |
Collapse
|
4
|
van den Besselaar AMHP, Cobbaert CM. Assignment of international normalized ratio to frozen and freeze-dried pooled plasmas. Clin Chem Lab Med 2020; 58:2089-2097. [PMID: 32840098 DOI: 10.1515/cclm-2019-1321] [Citation(s) in RCA: 2] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 12/23/2019] [Accepted: 03/29/2020] [Indexed: 11/15/2022]
Abstract
Objectives Frozen and freeze-dried plasmas may be used for local prothrombin time system calibration, for direct international normalized ratio (INR) determination, and for quality assessment. The purpose of the present study was to evaluate the usefulness of INRs assigned with various types of thromboplastins to frozen and freeze-dried pooled plasmas obtained from patients treated with vitamin K antagonists. Methods INRs were calculated according to the international sensitivity index (ISI) model using various thromboplastins and instruments, i.e. International Standards for thromboplastin as well as six commercial reagents prepared from rabbit and bovine brain, and recombinant human tissue factor. The uncertainty of the INRs was assessed using the standard deviations of clotting times and ISI values. Commutability of the plasmas was assessed according to the approved Clinical and Laboratory Standards Institute (CLSI) Guideline EP30-A. Validation of a set of six frozen plasma pools for direct INR determination was performed according to the Subcommittee on Control of Anticoagulation of the Scientific and Standardization Committee of the International Society on Thrombosis and Haemostasis (SSC/ISTH) guidelines. Results For all frozen and freeze-dried plasmas, the INRs calculated with bovine thromboplastin Thrombotest were lower than the INRs assigned with other thromboplastins. With a few exceptions, the frozen and freeze-dried pooled plasmas were commutable. When the set of six frozen plasma pools was used for local calibration, the analytical bias of the INR was less than ±10% for all commercial reagents except Thrombotest. Conclusions Processing of fresh plasmas to prepare pooled frozen plasmas and freeze-dried plasmas may lead to different INR assignments depending on the thromboplastin used. Despite minor INR differences, a set of six frozen plasma pools could be used for local calibration by direct INR determination.
Collapse
Affiliation(s)
- Antonius M H P van den Besselaar
- Coagulation Reference Laboratory, Department of Clinical Chemistry and Laboratory Medicine, Leiden University Medical Center, P.O. Box 9600, 2300 RC Leiden, The Netherlands.,Department of Thrombosis and Hemostasis, Leiden University Medical Center, Leiden, The Netherlands, Phone: +31 71 526 1942
| | - Christa M Cobbaert
- Coagulation Reference Laboratory, Department of Clinical Chemistry and Laboratory Medicine, Leiden University Medical Center, Leiden, The Netherlands
| |
Collapse
|
5
|
van den Besselaar AMHP, van Rijn CJJ, Abdoel CF, Chantarangkul V, Scalambrino E, Kitchen S, Tripodi A, Woolley AM, Padovan L, Cobbaert CM. Paving the way for establishing a reference measurement system for standardization of plasma prothrombin time: Harmonizing the manual tilt tube method. J Thromb Haemost 2020; 18:1986-1994. [PMID: 32356308 PMCID: PMC7496835 DOI: 10.1111/jth.14873] [Citation(s) in RCA: 8] [Impact Index Per Article: 2.0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 02/17/2020] [Revised: 04/20/2020] [Accepted: 04/22/2020] [Indexed: 11/30/2022]
Abstract
BACKGROUND International normalized ratio (INR) is traceable to World Health Organization (WHO) International Standards for thromboplastins. International Standards must be used with a manual tilt tube technique (MTT) for prothrombin time (PT) determination. An important part of the total variability of INR is due to poor harmonization of MTT across WHO reference laboratories. OBJECTIVES To determine the origins of PT differences between operators performing MTT and to develop a harmonized MTT. METHODS Two workshops were held where WHO reference laboratory operators could compare their PTs using MTT and the same equipment. A harmonized MTT was used by seven operators in the second workshop. RESULTS Differences have been observed in tilting frequency and in the height of pipetting plasma in the test tube. At the beginning of the first workshop, the tilting cycle time varied between 1.1 and 2.7 seconds. The mean PT of normal plasma obtained by pipetting plasma at the top of the tube was 14.3 seconds but was 12.9 seconds when plasma was pipetted at the bottom of the tube. When using the harmonized MTT for WHO International Standard rTF/16, the differences between operators were not greater than 1.1 seconds in normal plasma, and not greater than 1.3 seconds in patient plasma with average INR of 3.0. INR between-operator coefficient of variation was 2.3%. CONCLUSION Application of a harmonized MTT in three reference laboratories resulted in substantial reduction of between-operator variation of PT and INR. The harmonized MTT is proposed as Candidate Reference Measurement Procedure.
Collapse
Affiliation(s)
- Antonius M. H. P. van den Besselaar
- Coagulation Reference LaboratoryDepartment of Clinical Chemistry and Laboratory MedicineLeiden University Medical CentreLeidenThe Netherlands
| | - Claudia J. J. van Rijn
- Coagulation Reference LaboratoryDepartment of Clinical Chemistry and Laboratory MedicineLeiden University Medical CentreLeidenThe Netherlands
| | - Charmane F. Abdoel
- Coagulation Reference LaboratoryDepartment of Clinical Chemistry and Laboratory MedicineLeiden University Medical CentreLeidenThe Netherlands
| | - Veena Chantarangkul
- Angelo Bianchi Bonomi Hemophilia and Thrombosis CenterFondazione IRCCS Ca’ Granda Ospedale Maggiore Policlinico and Fondazione Luigi VillaMilanoItaly
| | - Erica Scalambrino
- Angelo Bianchi Bonomi Hemophilia and Thrombosis CenterFondazione IRCCS Ca’ Granda Ospedale Maggiore Policlinico and Fondazione Luigi VillaMilanoItaly
| | - Steve Kitchen
- Department of CoagulationRoyal Hallamshire HospitalSheffieldUK
| | - Armando Tripodi
- Angelo Bianchi Bonomi Hemophilia and Thrombosis CenterFondazione IRCCS Ca’ Granda Ospedale Maggiore Policlinico and Fondazione Luigi VillaMilanoItaly
| | | | - Lidia Padovan
- Angelo Bianchi Bonomi Hemophilia and Thrombosis CenterFondazione IRCCS Ca’ Granda Ospedale Maggiore Policlinico and Fondazione Luigi VillaMilanoItaly
| | - Christa M. Cobbaert
- Coagulation Reference LaboratoryDepartment of Clinical Chemistry and Laboratory MedicineLeiden University Medical CentreLeidenThe Netherlands
| |
Collapse
|