Kim HJ, Lee HK, Cho JH, Yang HJ. Quantitative comparison of transient elastography (TE), shear wave elastography (SWE) and liver biopsy results of patients with chronic liver disease.
J Phys Ther Sci 2015;
27:2465-8. [PMID:
26357427 PMCID:
PMC4563291 DOI:
10.1589/jpts.27.2465]
[Citation(s) in RCA: 13] [Impact Index Per Article: 1.4] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 03/19/2015] [Accepted: 04/24/2015] [Indexed: 12/12/2022] Open
Abstract
[Purpose] The purpose of this study was to carry out a comparitive analysis of hepatic
fibrosis results of the liver hardness of patients with chronic liver disease as measured
by elastography (TE), shear wave elastography (SWE), and liver biopsy. [Subjects and
Methods] This study was a retrospective analysis of 304 patients who underwent SWE and TE
before and after liver biopsy, taken from among patients who had been checked for liver
fibrosis by liver biopsy between August 2013 and August 2014. We used receiver operating
characteristic (ROC) curve to prove the diagnostic significance of liver stiffness, and
then analyzed the sensitivity, specificity, accuracy, positive predictive value, and
negative predictive value of SWE and TE, as well as the kappa index through cross-analysis
of SWE, TE, and liver biopsy. [Results] For liver hardness, the sensitivity of SWE was
84.39%, the specificity of SWE was 97.92%, the accuracy of SWE was 87.33%, the positive
predictive value of SWE was 99.32%, and the negative predictive value of SWE was 63.51%.
The sensitivity of TE was 94.80%, the specificity of TE was 77.08%, the accuracy of TE was
90.95%, the positive predictive value of TE was 93.97%, and the negative predictive value
of TE was 80.43%. [Conclusion] It is our opinion that SWE and TE are non-invasive methods
that are more effective than the invasive methods used for diagnosing liver hardness.
Invasive methods cover only a section of liver tissue, and are more likely to cause side
effects during biopsy.
Collapse