Ahmed T, Rahman NA, Alam MK. Validation and reliability of a prototype orthodontic bracket debonding device equipped with force-sensitive resistor (FSR): a novel method of measuring orthodontic bracket debonding force in vivo.
Prog Orthod 2019;
20:26. [PMID:
31281954 PMCID:
PMC6612523 DOI:
10.1186/s40510-019-0277-x]
[Citation(s) in RCA: 4] [Impact Index Per Article: 0.8] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 04/23/2019] [Accepted: 05/23/2019] [Indexed: 12/01/2022] Open
Abstract
Background
To introduce an orthodontic bracket debonding device capable of measuring debonding force clinically by a novel sensor mechanism
Materials and method
A prototype orthodontic debonding device was constructed utilizing a lift-off debonding instrument (LODI) and force-sensitive resistor (FSR). For data interpretation, the force sensor was equipped with a microcontroller and C++ programming software running on a computer. Ninety-nine (99) 0.022-in. conventional metallic brackets were bonded to premolar teeth in vitro by a single clinician applying the same adhesive and bonding technique. For validation, the mean debonding force measured by the prototype debonding device (n = 30) and the universal testing machine (n = 30) was compared. Both intra- and inter-examiner reliability tests were done by holding and operating the device in a standardized manner. Following debonding by the prototype device, the bracket failure pattern was evaluated (n = 30) by adhesive remnant index (ARI) under the stereomicroscope at × 30 magnification. Statistical analysis included independent samples t test for validation and intraclass correlation coefficient (ICC) with a 95% confidence interval for both intra- and inter-examiner reliability.
Results
Mean orthodontic bracket debonding force measured by the prototype device (9.36 ± 1.65 N) and the universal testing machine (10.43 ± 2.71 N) was not significantly different (p < 0.05). The prototype device exhibited excellent intra- [ICC (3, 1) = 0.942] and inter-examiner reliability [ICC (2, 1) = 0.921] and was able to debond brackets mostly at the bracket-adhesive interface.
Limitation
Due to adjusting the position and mechanism of the force sensor, the device had to be held in a modified standardized position.
Conclusion
A novel method of measuring in vivo orthodontic bracket debonding force has been introduced which proved to be validated, reliable, and safe in terms of enamel damage.