Ueda M, Saito T, Fujimori K, Moriuchi T, Shibutani Y. Potentiometric Performance of Silver Ion-Selective Electrodes Based on Tridentate Schiff Base Derivatives.
ANAL SCI 2004;
20:1649-53. [PMID:
15636510 DOI:
10.2116/analsci.20.1649]
[Citation(s) in RCA: 11] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/23/2022]
Abstract
To examine the directivity for improving the silver ion discrimination ability of a Schiff base, three kinds of tridentate ligands were synthesized and compared with the similar quadridentate ligand as the silver ionophore. Among the Schiff base derivatives tested, 3-(2-pyridylethylimino)-2-butanoneoxime, having one oxime and a pyridine substituent, was found to be the best ionophore for a silver-ion electrode. The electrode based on this derivative exhibited good silver-ion selectivity, -log Kpot(Ag+,K+) = 3.8, comparable to that of a quadridentate Schiff base, N,N'-bis(2'-hydroxyimino-1'-phenylpropyleden)-1,3-propanediamine, reported previously, except for a pseudo Nernstian response (35.6 mV decade(-1)) with a wide silver-ion activity change in the activity change from 5.0 x 10(-7) to 7.9 x 10(-2) mol dm(-3).