Wood NA, Schwartzman D, Passineau MJ, Moraca RJ, Zenati MA, Riviere CN. Beating-heart registration for organ-mounted robots.
Int J Med Robot 2018;
14:e1905. [PMID:
29508506 DOI:
10.1002/rcs.1905]
[Citation(s) in RCA: 2] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 04/22/2017] [Revised: 01/23/2018] [Accepted: 01/30/2018] [Indexed: 12/25/2022]
Abstract
BACKGROUND
Organ-mounted robots address the problem of beating-heart surgery by adhering to the heart, passively providing a platform that approaches zero relative motion. Because of the quasi-periodic deformation of the heart due to heartbeat and respiration, registration must address not only spatial registration but also temporal registration.
METHODS
Motion data were collected in the porcine model in vivo (N = 6). Fourier series models of heart motion were developed. By comparing registrations generated using an iterative closest-point approach at different phases of respiration, the phase corresponding to minimum registration distance is identified.
RESULTS
The spatiotemporal registration technique presented here reduces registration error by an average of 4.2 mm over the 6 trials, in comparison with a more simplistic static registration that merely averages out the physiological motion.
CONCLUSIONS
An empirical metric for spatiotemporal registration of organ-mounted robots is defined and demonstrated using data from animal models in vivo.
Collapse