Lee KH, Kwon JW, Yoon YC, Choi SH, Jung JY, Kim JH, Lee SJ. Slot-scan digital radiography of the lower extremities: a comparison to computed radiography with respect to image quality and radiation dose.
Korean J Radiol 2009;
10:51-7. [PMID:
19182503 PMCID:
PMC2647173 DOI:
10.3348/kjr.2009.10.1.51]
[Citation(s) in RCA: 6] [Impact Index Per Article: 0.4] [Reference Citation Analysis] [Abstract] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/20/2022] Open
Abstract
Objective
To compare the slot-scan digital radiography (SSDR) of the lower extremity region and the computed radiography (CR) method with respect to the image quality and radiation exposure.
Materials and Methods
We enrolled 54 patients who underwent both the SSDR and CR of the lower extremities. The study evaluated and statistically compared the image quality of four features (outer cortex, inner cortex, trabeculae and intermuscular fat) at six different levels (pelvis, hip, femur, knee, tibia and ankle) between each method. The image quality was evaluated using a visibility scale, and the entrance skin dose was measured using a dosimeter at three different levels of a phantom (hip, knee, and ankle).
Results
The mean image visibility scale values for the SSDR method were significantly higher than for the CR method. The entrance skin dose for the SSDR method was 278 µGy at each level, compared to the entrance skin doses of the CR method, which were 3,410 µGy for the hip, 1,152 µGy for the knee, and 580 µGy for the ankle.
Conclusion
Both the image quality and patient entrance skin dose data suggest that the SSDR method is superior to the CR method for the lower extremity musculoskeletal examination.
Collapse