Rogati G, Caravaggi P, Leardini A, Erani P, Fognani R, Saccon G, Boriani L, Baleani M. A novel apparatus to assess the mechanical properties of Ankle-Foot Orthoses: Stiffness analysis of the Codivilla spring.
J Biomech 2022;
142:111239. [PMID:
35940017 DOI:
10.1016/j.jbiomech.2022.111239]
[Citation(s) in RCA: 2] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 05/23/2022] [Revised: 07/13/2022] [Accepted: 07/29/2022] [Indexed: 10/16/2022]
Abstract
Ankle-Foot Orthoses (AFOs) are the most common devices prescribed to support the ankle and restore a quasi-normal gait pattern in drop-foot patients. AFO stiffness is possibly the main mechanical property affecting foot and ankle biomechanics. A variety of methods to evaluate this property have been reported, however no standard procedure has been validated and widely used. This study is reporting the repeatability of a novel apparatus to measure AFO stiffness in ideal frictionless conditions. The apparatus is based on a servo-hydraulic testing machine and allows to apply a displacement-controlled rotation of the AFO shell, simulating the physiological ankle dorsi/plantarflexion movement. The repeatability of the apparatus in measuring AFO stiffness in dorsiflexion and plantarflexion was assessed intra- and inter-session in a sample of standard polypropylene AFOs of different sizes (Codivilla spring). The repeatability of the apparatus in measuring the AFO stiffness was high. The Intra- and Inter-session Coefficient of Variation ranged between 0.02 ÷ 1.3 % and 1.3 ÷ 5 %, respectively. The Intra Class Correlation Coefficient ranged between 0.999 ÷ 1 intra- and 0.993 ÷ 0.997 inter-session. AFOs stiffness was observed to increase with the AFO size. The setup is easy to replicate and can be implemented with any torsion-controlled servo-hydraulic testing machine and has resulted simple to use and flexible enough to adapt to AFOs with different sizes. The frictionless contacts characterizing the apparatus make it possible to measure the ideal AFO stiffness by excluding the effect of the fixation methods to the leg and help to improve the repeatability of measurements.
Collapse