Aramburú JS, Gehrke SA, Dedavid BA, Treichel TLE, Pippi NL. Correlation of Fracture Resistance of Dental Implants and Bite Force in Dogs described in the literature: An
In Vitro Study.
J Vet Dent 2021;
38:75-80. [PMID:
34736337 DOI:
10.1177/08987564211044940]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/17/2022]
Abstract
Dental implants are not routinely used for rehabilitation in veterinary dentistry. For some veterinarians, further studies are necessary to be considered for clinical use in animals. The objective of the present in vitro study was to evaluate static fatigue of dental implants and to correlate that with the bite strength of dogs described in the literature. Sixty implants and abutments were used with the smallest diameter of each brand of implant utilized in the study. Three groups (n = 20) were created on the basis of the implant diameter, all with external hex connector: 3.30 mm (group 1), 4.0 mm (group 2) and 5.0 mm (group 3). All groups were subjected to quasi-static loading at 30° to the implant's long axis in a universal machine (model AME-5 kN). The mean fracture strength for group 1 was 964 ± 187 N, for group 2 was 1618 ± 149 N and for group 3 was 2595 ± 161 N. Significant differences between the groups with respect to resistance after the load applications were observed (P < .05). The diameter of implants affects the resistance to external forces during the application of non-axial strength (off-axis loading) and must be considered during the planning of rehabilitation to avoid problems.
Collapse