Cernicanu A, Lepetit-Coiffe M, Roland J, Becker CD, Terraz S. Validation of fast MR thermometry at 1.5 T with gradient-echo echo planar imaging sequences: phantom and clinical feasibility studies.
NMR IN BIOMEDICINE 2008;
21:849-858. [PMID:
18574794 DOI:
10.1002/nbm.1267]
[Citation(s) in RCA: 35] [Impact Index Per Article: 2.2] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 05/26/2023]
Abstract
The purpose of this work was to validate in phantom studies and demonstrate the clinical feasibility of MR proton resonance frequency thermometry at 1.5 T with segmented gradient-echo echo planar imaging (GRE-EPI) sequences during liver tumour radiofrequency (RF) ablation. Classical GRE acquisitions and segmented GRE-EPI acquisitions were performed at 1.5 T during simultaneous RF heating with an MR-compatible RF electrode placed in an agar gel phantom. Temperature increments were calculated and compared with four optical temperature probe measurements using Bland- Altman analysis. In a preliminary clinical feasibility study, the rapid GRE-EPI sequence (echo train length = 13) was used for MR temperature monitoring of RF ablation of liver tumours in three patient procedures. For phantom experiments, the Bland-Altman mean of differences between MR and optical probe temperature measurements was <0.4 degrees C, and the 95% limits of agreement value was <1.4 degrees C. For the in vivo studies, respiratory-triggered GRE-EPI acquisitions yielded a temperature accuracy of 1.3 +/- 0.4 degrees C (acquisition time = 0.6 s/image, spatial coverage of three slices/respiratory cycle). MR proton resonance frequency thermometry at 1.5 T yields precise and accurate measurements of temperature increment with both classical GRE and rapid GRE-EPI sequences. Rapid GRE-EPI sequences minimize intra-scan motion effects and can be used for MR thermometry during RF ablation in moving organs.
Collapse