Hoffmann JJML. Basophil counting in hematology analyzers: time to discontinue?
Clin Chem Lab Med 2020;
59:cclm-2020-1528. [PMID:
33554563 DOI:
10.1515/cclm-2020-1528]
[Citation(s) in RCA: 4] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 10/14/2020] [Accepted: 12/02/2020] [Indexed: 02/24/2024]
Abstract
Basophils (basophilic granulocytes) are the least abundant cells in blood. Nowadays, basophils are included in the complete blood count performed by hematology analyzers and therefore reported in practically all patients in whom hematologic investigations are requested. However, hematology analyzers are not reliable enough to report clinically useful results. This is due to a combination of very high analytical imprecision and poor specificity, because the chemical and physical methods used for basophil counting in hematology analyzers are ill-defined and thus basophils are not well recognized by the analyzers. As a result, false basophil counts are quite common. In view of increasing analytical performance demands, hematology laboratories should stop reporting basophil counts produced by hematology analyzers. Suggestions for alternative pathways are presented for those situations where basophils are of clinical relevance.
Collapse