Dolotov LE, Sinichkin YP, Tuchin VV, Utz SR, Altshuler GB, Yaroslavsky IV. Design and evaluation of a novel portable erythema-melanin-meter.
Lasers Surg Med 2004;
34:127-35. [PMID:
15004824 DOI:
10.1002/lsm.10233]
[Citation(s) in RCA: 41] [Impact Index Per Article: 2.0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/10/2022]
Abstract
BACKGROUND AND OBJECTIVES
Objective evaluation of the pigmentation index (PI) and the eythema index (EI) of human skin is a prerequisite for successful optimization of laser- and intense-pulsed-light (IPL)-based treatment modalities in dermatology.
STUDY DESIGN/MATERIALS AND METHODS
We describe a three-wavelength technique for determining PI and EI as well as its particular implementation using LEDs operating at wavelengths of 560, 650, and 710 nm and a large-area photodiode. The instrument has been evaluated both in vitro and in vivo.
RESULTS
In vitro, good correlation between the measured indices and results obtained with commercially available techniques has been observed. In addition, linearity of the PI with melanin concentration in the phantom medium up to 7 x 10(-3) nm(-1) (defined as a slope of the optical density spectrum) has been established. In vivo, feasibility of using the technique for predicting the minimal erythema dose (MED), minimal phototoxic dose (MPD), and the threshold of epidermal damage in a photothermal treatment has been demonstrated.
CONCLUSIONS
The data suggest that the technique has a substantial potential as a method of pre-treatment diagnostics for photochemical and photothermal procedures.
Collapse