van der Beek FB, Briaire JJ, Frijns JHM. Population-based prediction of fitting levels for individual cochlear implant recipients.
Audiol Neurootol 2014;
20:1-16. [PMID:
25413720 DOI:
10.1159/000362779]
[Citation(s) in RCA: 14] [Impact Index Per Article: 1.4] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 10/11/2012] [Accepted: 04/09/2014] [Indexed: 11/19/2022] Open
Abstract
OBJECTIVES
This study analyzed the predictability of fitting levels for cochlear implant recipients based on a review of the clinical levels of the recipients.
DESIGN
Data containing threshold levels (T-levels) and maximum comfort levels (M-levels) for 151 adult subjects using a CII/HiRes 90K cochlear implant with a HiFocus 1/1 J electrode were used. The 10th, 25th, 50th, 75th and 90th percentiles of the T- and M-levels are reported. Speech perception of the subjects, using a HiRes speech coding strategy, was measured during routine clinical follow-up.
RESULTS
T-levels for most subjects were between 20 and 35% of their M-levels and were rarely (<1/50) below 10% of the M-levels. Furthermore, both T- and M-levels showed an increase over the first year of follow-up. Interestingly, levels expressed in linear charge units showed a clear increase in dynamic range (DR) over 1 year (29.8 CU; SD 73.0), whereas the DR expressed in decibels remained stable. T-level and DR were the only fitting parameters for which a significant correlation with speech perception (r = 0.34, p < 0.01, and r = 0.33, p < 0.01, respectively) could be demonstrated. Additionally, analysis showed that T- and M-level profiles expressed in decibels were independent of the subjects' across-site mean levels. Using mixed linear models, predictive models were obtained for the T- and M-levels of all separate electrode contacts.
CONCLUSIONS
On the basis of the data set from 151 subjects, clinically applicable predictive models for T- and M-levels have been obtained. Based on one psychophysical measurement and a population-based T- or M-level profile, individual recipients' T- and M-levels can be approximated with a closed-set formula. Additionally, the analyzed fitting level data can serve as a reference for future patients.