Ketata H, Hfaiedh N, Kanhonou M, Badreddine H. Impact of fractured tibia implant fixation devices on bone stiffness during bending test.
Med Eng Phys 2024;
131:104228. [PMID:
39284654 DOI:
10.1016/j.medengphy.2024.104228]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 12/15/2023] [Revised: 07/05/2024] [Accepted: 08/13/2024] [Indexed: 09/19/2024]
Abstract
This study focuses on evaluating the failure resistance of a previously reduced tibia with internal fixation implants as PLate (PL) or InterMedullary Nail (IMN), subjected later to a tibial lateral trauma. To replicate this type of trauma, which can be caused by a road accident, a three-point bending test is considered using experimental tests and numerical simulations. The withstand evaluation of the tibia-PL and tibia-IMN structures was conducted by following the load transfer through, the bone and the used implants. The analysis, up to tibia failure, required the use of an elasto-plastic behavior law coupled to damage. The model parameters were identified using experimental tests. Il was shown that the tibia-IMN structure provided a bending resistant load up to three-times higher than the tibia-PL. In fact, the used screws for plate fixation induced a high level of stress in the vicinity of threaded region, leading to a crack initiation and a damage propagation. However, in tibia-IMN structure the highest stress was generated in the trapped zone between the loader and the nail, promoting crack formation. From a biomechanical point of view, the structure with IMN is safer than the structure with PL, whose fixation induces earlier damage in bone.
Collapse