Abstract
OBJECTIVE
The purpose of the study was to evaluate the precision, accuracy, and safety of three different modes of excimer laser surgery for myopic astigmatism.
DESIGN
A prospective interventional case series.
PARTICIPANTS
A total of 150 eyes with compound myopic astigmatism were treated.
INTERVENTION
Ablation programs included the VISX Twenty/Twenty excimer laser using either the sequential mode (58 eyes) or the elliptical mode (54 eyes) and the Technolas-Chiron Keracor 116 hybrid scanning laser (38 eyes).
MAIN OUTCOME MEASURES
Reduction in cylinder and sphere, vector analysis (polar values), uncorrected visual acuity, and best spectacle-corrected visual acuity were evaluated at 1, 6, and 12 months.
RESULTS
By 12 months, sphere was reduced by 95.1%, 87.7%, and 75% in low-astigmatism eyes treated with the VISX sequential, VISX elliptical, and Technolas modes, respectively, and by 92.2%, 98%, and 77.6% in high-astigmatism eyes. By 12 months, cylinder was reduced by 51.5%, 72.2%, and 36.3% in low-astigmatism eyes and by 70%, 78%, and 45.6%, respectively, in high-astigmatism eyes. Vector analysis showed that the elliptical treatment produced more significant reductions in mean polar values than the other two treatments at various timepoints.
CONCLUSION
These results suggest that the ablation approach may influence the clinical result.
Collapse