Sanders DR. Matched Population Comparison of the Visian Implantable Collamer Lens and Standard LASIK for Myopia of -3.00 to -7.88 Diopters.
J Refract Surg 2007;
23:537-53. [PMID:
17598571 DOI:
10.3928/1081-597x-20070601-02]
[Citation(s) in RCA: 59] [Impact Index Per Article: 3.5] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/20/2022]
Abstract
PURPOSE
To compare matched populations of LASIK and Visian Implantable Collamer Lens (ICL) cases in the correction of myopia between -3.00 and -7.88 diopters (D).
METHODS
One hundred sixty-four LASIK eyes with prospective data collected from a single center and 164 ICL eyes from the multicenter US ICL Clinical Trial were compared in this observational non-randomized study. The LASIK and ICL groups were well matched for age, gender, and mean level of preoperative spherical equivalent refraction.
RESULTS
At 6 months, best spectacle-corrected visual acuity (BSCVA) > or = 20/20 was 85% with LASIK and 95% with ICL (P = .003) compared to preoperative values of 93% and 88%, respectively (P = .292). Loss of > or = 2 lines of BSCVA was significantly lower with the ICL at 1 week (0.6% vs 10%, P < .001) and 1 month (7% vs 0%, P = .001) with comparable outcomes at 6 months (0% vs 1%). At 6 months postoperatively, uncorrected visual acuity (UCVA) > or = 20/15 (11% vs 25%, P = .001) and > or = 20/20 (49% vs 63%, P = .001) was better in the ICL cases. Predictability within 0.50 D at 6 months for ICL cases was 85% (67% LASIK, P < .001); 97% of ICL cases were within 1.00 D (88% LASIK, P = .002). Refractive stability (+/- 0.50 D) between 1 and 6 months was 93% with ICL compared to only 82% with LASIK (P = .006).
CONCLUSIONS
The ICL performed better than LASIK in almost all measures of safety, efficacy, predictability, and stability in this matched population comparison, supporting the ICL as an effective alternative to existing refractive laser surgical treatments for the range of myopia studied.